User's Manual
Table Of Contents
- Introducing IntelliVue Telemetry
- Product Safety
- Transceiver Controls
- Basic Operation
- Alarms
- ECG Monitoring
- ST/AR Arrhythmia Monitoring
- SpO2 Monitoring
- Telemetry Functions at the Information Center & TeleMon
- Telemetry Functions at the Information Center
- Telemetry Controls in the Patient Window
- Locating the Transceiver (Find Device)
- Viewing Device Location in the Patient Window (optional)
- Viewing Device Location History (optional)
- Using the Device Location Client (optional)
- Patient Configurable Settings in Telemetry Setup
- Unit- Configurable Settings
- RF Auto Shutoff
- Transceiver Operation when Connected to TeleMon
- Telemetry Functions at the Information Center
- Pairing Monitoring Devices
- Maintenance, Cleaning & Troubleshooting
- Safety Standards & Specifications
- Accessories
- Sales and Support Offices
Battery Information
4-18 Basic Operation
Battery Levels
Battery Gauge
Approximate
Battery Life*
Remaining
Approximate
Operating
Time*
Remaining
Functionality
4 green indicators > 75% > 34.7 hours Normal operation
3 green indicators > 50% > 23.1 hours Normal operation
2 green indicators > 25% > 11.6 hours Normal operation
1 green indicator 25% to Battery
Low level
> 15 minutes Normal operation
1 red indicator Battery Low
level to
Replace
Battery level
< 15 minutes Normal operation
(see note)
no indicator Replace
Battery level
(Check
batteries for
correct
polarity)
none Transceiver
shutdown/RF
shutoff
* Battery life times are based on use of Duracell MN 1500 batteries in a TRx
transceiver not equipped with a Short-Range Radio Adapter. Times for TRx
+
are somewhat lower. Life times when other battery brands are used may be
different.
Note—For the TRx
+
transceiver, SpO
2
functionality is disabled when the red
light (Battery Low) is lit.