Physiologic Alarms
5-4 Alarms
Unsuspending
& Resuming
Alarms
Alarms will be resumed automatically after the configured suspend duration
time. You can cancel alarm suspend manually before the configured suspend
duration time has expired from the transceiver (see following directions) or from
the Information Center.
Physiologic Alarms
Physiologic alarms indicate a life-threatening situation or a less urgent situation
such as heart rate beyond limits. There are no physiologic alarm signals
generated by the transceiver.
All physiologic alarms are generated at the IntelliVue Information Center,
and all alarm signals must be acknowledged at the Information Center or
an assigned IntelliVue Patient Monitor.
Step Action
1Press the
Telemetry and Check buttons simultaneously until the
indicator turns off.
The Alarm Suspend icon is turned off.
The message "Alarms Suspended" (or "Alarms Paused") is
removed from the Information Center display.
2 Alternately, you can unsuspend alarms at the Information Center.
See IntelliVue Information Center Instructions for Use or Online
Help for directions.
For instructions on how to resume alarms from TeleMon, see the
TeleMon Instructions for Use.