Physiologic Alarms
Alarms
5-7
** SpO
2
T yyy > xxx Yellow High SpO
2
.
SpO
2
value (yyy) greater than high SpO
2
limit (xxx).
SpO
2
** SpO
2
T yyy < xxxx Yellow Low SpO
2
.
SpO
2
value (yyy) less than low SpO
2
limit
(xxx).
SpO
2
** ST lead > xxx Yellow STlead is greater than the Elevation limit. ST/AR
** ST lead < xxx Yellow STlead is less than the Depression limit. ST/AR
* HR yyy > xxx Yellow Heart Rate (yyy) greater than the upper
Heart rate limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
* HR yyy < xxx Yellow Heart Rate (yyy) lower than the lower
Heart Rate limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
* IRREGULAR HR Yellow Consistently irregular rhythm (irregular
R-R intervals).
ST/AR
Enhanced
Arrhythmia
* MISSED BEAT Yellow No beat detected for 1.75 x average R-R
interval for Heart Rate greater than 120,
or no beat for 1 second with Heart Rate
greater than 120 (non-paced patient only).
ST/AR
Enhanced
Arrhythmia
* MULTIFORM PVCs Yellow The occurrence of two differently shaped
Vs, each occurring at least twice within
the last 300 beats as well as each
occurring at least once within the last 60
beats.
ST/AR
Enhanced
Arrhythmia
Physiologic (Patient) Alarms
Alarm Text Priority Condition Source