Technical Alarms (INOPs)
5-12 Alarms
CENTRAL: TELE ONLY
Source - Patient Monitor
Hard System connectivity via the
transceiver is limited (no
alarms, only local
numerics) when in
companion mode and host
monitor does not have
system connectivity. Only
the transceiver’s
measurements and alarms
can be displayed at the
Information Center.
If the disconnection of the
host monitor is not
intentional, re-connect the
monitor to the network.
CHECK ECG SETTINGS
Source - Telemetry
Hard Synchronization of ECG
settings between the
monitor and the
Information Center has
failed.
Check that the ECG settings
from the ECG source are
appropriate.
Note—When transitioning
between networked and non-
networked monitoring, this
INOP will display. Pressing
the Silence button will
dismiss the INOP.
CHECK ECG SOURCE
Source - Patient Monitor
Yellow
Technical
Alarm
The transceiver and the
monitor both have valid
ECG signals.
Unpair the transceiver and
monitor or remove the
leadset from the device that
is no longer the ECG source.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do