User's Manual

Table Of Contents
Technical Alarms (INOPs)
5-18 Alarms
SpO
2
T NOISY SIGNAL
Source - Telemetry
Hard Excessive patient
movements or electrical
interference are causing
irregular pulse patterns
Reduce movement or
electrical noise sources.
SpO
2
T NON-PULSATILE
Note—When paired
directly with an IntelliVue
MP5 Patient Monitor, the
INOP will display as
SpO
2
T SENSOR OFF.
Source - Telemetry
Hard Pulse is too weak or not
detectable
Check connection to patient.
Change sensor site. Avoid
site distal to BP cuff or intra-
arterial line.
SpO
2
T SENSOR MALF
Source - Telemetry
Hard Malfunction of the SpO
2
sensor/adapter cable
Replace sensor.
TAAP DISABLED
Source - Patient Monitor
Hard The monitor’s telemetry
configuration does not
allow connection of the
transceiver to the monitor.
Check configuration settings
at the monitor.
Contact Service.
TELE CONFIG
UNSUPPORTED
Source - Patient Monitor
Hard Transceiver not supported
(companion mode).
Make sure the transceiver is
the correct revision.
Contact Service.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do