Technical Alarms (INOPs)
5-20 Alarms
TELE UNSUPPORTED
Source - Patient Monitor
Hard The transceiver is not
supported for direct
connection to the monitor.
Contact Service. Make sure
the transceiver is the correct
revision.
TELE WEAK SIGNAL
Source - Telemetry
Soft Patient is at outer
range of the radio
coverage area.
Telemetry pack is
receiving a weak
signal with high data
loss from the AP.
Condition exists for
multiple devices in a
specific area
Return patient to the
coverage area.
If patient is in close
proximity to AP,
replace telemetry pack.
Contact service.
The AP covering the
specific area is
suspect. Contact
Service.
TRANSMITTER MALF
Source - Telemetry
Hard Transceiver malfunction Contact Service to replace
the transceiver.
TRANSMITTER OFF
Source - Telemetry
Hard RF Auto Shutoff after 10
minutes of all leads off and
no SpO
2
sensor connected.
Reattach ECG leads to
patient.
Reattach SpO
2
sensor.
Press the Check
button.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do