User's Manual
Table Of Contents
- Introducing IntelliVue Telemetry
- Product Safety
- Transceiver Controls
- Basic Operation
- Alarms
- ECG Monitoring
- ST/AR Arrhythmia Monitoring
- SpO2 Monitoring
- Telemetry Functions at the Information Center & TeleMon
- Telemetry Functions at the Information Center
- Telemetry Controls in the Patient Window
- Locating the Transceiver (Find Device)
- Viewing Device Location in the Patient Window (optional)
- Viewing Device Location History (optional)
- Using the Device Location Client (optional)
- Patient Configurable Settings in Telemetry Setup
- Unit- Configurable Settings
- RF Auto Shutoff
- Transceiver Operation when Connected to TeleMon
- Telemetry Functions at the Information Center
- Pairing Monitoring Devices
- Maintenance, Cleaning & Troubleshooting
- Safety Standards & Specifications
- Accessories
- Sales and Support Offices
Technical Alarms (INOPs)
5-20 Alarms
TELE UNSUPPORTED
Source - Patient Monitor
Hard The transceiver is not
supported for direct
connection to the monitor.
Contact Service. Make sure
the transceiver is the correct
revision.
TELE WEAK SIGNAL
Source - Telemetry
Soft • Patient is at outer
range of the radio
coverage area.
• Telemetry pack is
receiving a weak
signal with high data
loss from the AP.
• Condition exists for
multiple devices in a
specific area
• Return patient to the
coverage area.
• If patient is in close
proximity to AP,
replace telemetry pack.
Contact service.
• The AP covering the
specific area is
suspect. Contact
Service.
TRANSMITTER MALF
Source - Telemetry
Hard Transceiver malfunction Contact Service to replace
the transceiver.
TRANSMITTER OFF
Source - Telemetry
Hard RF Auto Shutoff after 10
minutes of all leads off and
no SpO
2
sensor connected.
• Reattach ECG leads to
patient.
• Reattach SpO
2
sensor.
• Press the Check
button.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do