4 HP Dravidian Wave Viewer This chapter provides information about the HP Dravidian Wave Viewer, a clinical application run on an HP palmtop computer and used in conjunction with the HP Dravidian telemetry transmitter. For environmental specifications, see Appendix A. Topics include: = Installation of Wave Viewer m Exiting from Wave Viewer mn Configuration Screens w System Setup (These functions require a password.
Caution $ 3. wr Insert the wave viewer flash disk card into the left end of the palmtop. . Press to reboot the system. The “Welcome to the HP Dravidian Wave Viewer” screen will be displayed, followed by the “Communication Disrupted” screen. If the wave viewer does not start when the disk card is inserted, the palmtop may have insufficient memory. One (1) megabyte of memory is required to run the wave viewer, . To access patient waveform, connect the palmtop to the transmitter.
Configuration Screens The configuration screens enable you to view transmitter and wave viewer settings, and to access the Setup screens. Transmitter status information remains available even after the transmitter has been disconnected if another transmitter has not been connected. When the wave viewer is connected to a different transmitter, information for the initial transmitter will be erased. Dynamic status information is updated approximately once per second.
2. Select Setup. . Enter the password, followed by (ENTER). 4. Select Demo, then Yes. Be sure the palmtop is not near the transmitter. w Or to return to mealtime mode, press No Cancel . To Exit (Return to Mealtime Mode) 1. Select Exit Demo, then Exit Setup and Exit Con fig. Batteries HP supports only 1.5 volt, size AA alkaline or nickel-cadmium rechargeable batteries in the HP palmtop for use with the wave viewer application. Under typical usage, the life cycle of fresh alkaline batteries is 2-8 weeks.
Palmtop Power If there has been no key press on the palmtop for 10 Save Mode minutes, the palmtop will shut off automatically to conserve power. To resume operation, press (ON). Care and Cleaning For instructions on care and cleaning of the palmtop, refer to the HP palmtop documentation.
Inoperative Conditions This chapter gives you additional information about telemetry inoperative conditions (SYNOPSIS. A complete summary of In Ops and alarms can be found in Chapter 3 of the User's Guide. Monitoring During Inoperative conditions {IN OPS) exist when the signal Inoperative cannot be properly processed. IN OPS automatically i. additional s. Ever Condition reset when the condition clears. Every IN OP alarm is nations accompanied by a specific message.
5-2 Note $ Inoperative Conditions When fallback occurs, ECG-CH 2 occupies the sector. An IN OP message, for example LEADS OFF II, appears above ECG-Chi's waveform and the prompt Leads swapped due to lead off appears in the ECG Task Windows. When the condition clears, the IN OP message is removed and ECG-CH 1 returns to the sector. To prevent lead selection which would terminate the fallback mode, the soft key enabling the change of lead or label {or swap) is removed.
Fallback Mode with Arrhythmia Extended Monitoring Mode Note 9 Arrhythmia fallback is a feature of arrhythmia monitoring and occurs even if your telemetry system is not configured for fallback. Arrhythmia fallback occurs if your patient is arrhythmia monitored, ECG-CH 1 is IN OP, and ECG-CH 2 is operative. After a 10 second delay, ECG-CH 2 switches to the sector and arrhythmia analysis and alarms continue to be generated.
with ECG-CH 1 as Lead II and ECG-CH 2 as off, if the left leg (LL) electrode fall off, ECG-CH 2, Lead MCL, turns on, occupies the sector, and becomes the tachycardia lead. When the IN OP of ECG-CH 1 is corrected, monitoring continues as before. Extended Monitoring with the 4-wire Lead Set {(M1400A /B transmitter) Since the 4-wire lead set reconstructs four of its possible 6 leads, ITI, aVR, aVL, aVF, it may be able to continue monitoring if ECG-CH 1 and B are both IN OP.
Configuration Configuration How your telemetry system performs depends in large Choices part on the configuration choices made during system installation. The following tables summarize the factory-set defaults and the alternative configuration choices that relate to clinical practice. Configuration is performed at the receiver mainframe, except for the Dravidian transmitters, which are configured at the wave viewer, and all settings except frequency pertain to all receivers in the mainframe.
M2604A Receiver Mainframe Configuration Settings Item Factory Default User Choices GENERAL ALARM PARAMETERS Alarm Se spend 3 Minutes 3 Minutes, Infinite Slavic Tidewater ON ON, OFF GENERAL ECG PARAMETERS Palmer limits High 120 20-250 Low 50 15-2451 Tend Fallback ON ON, OFF Extended Monitoring ON ON, OFF Amigo SERIES TRANSMITTERS CG Bandwidth SH Monty CHE = Mo miter CH and CH2 = Monitoring, Exercise, Diagnostic, Paced, 51 GEOCACHE OFFER Lead Selection 4 Electrode CHUG TT CHINTZ HHL AVR YE aVF Cain t Cte 1 LL
M2604A Receiver Mainframe Configuration Settings (Cont.
For configuration of the following items, see HP Dravidian Telemetry System Installation and Configuration Guide: um Auto Self Test » Self-test Strip ® SDN Unit Number m SDN Branch Number m Country Code mn Locale Code m Frequencies HP Dravidian Transmitter Configuration Settings Rem Factory Dealt User Choices Dora Selection Bowie fend set Ma Yea.
a Changing the In general, configuration changes are best made by the Configuration service department. However, occasionally you may To Configure a Replacement HP Dravidian Transmitter Note 9 be called on to resolve a troublesome situation.
6-6 Configuration Transmitter Configuration Wave Viewer Lead Selection with 3-wire Lead Set: NO Ability for User to Change Frequency: YES Automatic Shutoff of Transmitter After 10 Minutes: YES Press Toff store settings. Era ET Er FIC . From the wave viewer Main Screen, select Con fig. oe Select Setup. . Enter the password, followed by (ENTER). . Select Chang Freq. ooh o . Enter the frequency for the replacement transmitter found on the central monitor, followed by (ENTER). f.
To Change the Frequency: Note 9 Hester figuration HF soulfulness ie wd ge thin enlighten Ene mien Flog HR an a Sardine nation we save he regurgitation by Sites Diet Mh attach Sh ow: Intelligentsia oes EF Toni _F2 FIO the central monitor, select Learn Code from the Telemetry Frequency screen {or in the Admit/Discharge window) for the highlighted bed and within 10 seconds, press on the replacement transmitter to enable the system to learn the new frequency.
e. Enter the new frequency for the receiver, followed by (ENTER). f. Enter the check code, followed by (ENTER). g. Press (Confirm) 2. Set the new frequency for the transmitter. a. From the wave viewer Main Screen, select Con fig. b. Select Setup. c. Enter the password, followed by (ENTER). d. Select Chang Freq. e. Enter the new frequency for the transmitter, followed by (ENTER). f. Enter the check code, followed by (ENTER). gTo set the new frequency, select Confirm Change Frequency Current Frequency. 422.
Installation and Patient Safety Caution $ Warning 9 Installation Information Environment Installation and setup must be performed by an HP service representative or designer. bad Power modules for analog output, antennas, and palmtop personal computers (wave viewer) must not be installed or used within an 8’ radius of the patient. Follow the instructions below to ensure a completely safe electrical installation.
7-2 Power Source Requirements Grounding the HP Dravidian Telemetry System Warning 9 Condensation Warning 9 circuits. Allow at least 2 inches (5 cms) clearance around the instrument (except the transmitter) for proper air circulation. The HP Dravidian Telemetry System can be operated from an AC source of 100/120/220/240 VAC delectable + 10%, with a frequency range The power consumption is 110 VA maximum, 95 VA average, 81 W maximum, 72 W average with 8 receiver modules.
Explanation of The following is an explanation of the symbols found on symbols the HP Dravidian Telemetry System: ~ v = Indicates the instrument is Type CF. (With define protection.) ° x Indicates the equipment is Type B. ~~ AC line current. fT Attention, consult accompanying documents. o> An electrical output. — An electrical input. fr Fuse input. J Grounding system. ! o Potentiality grounding system. @ Protective earth ground. &— Date in. @ Data In/Out. y Antenna input. J ~< Active Antenna Combiner.
7-4 These symbols indicate that the various instruments connected to the HP Dravidian Telemetry System are either Type B and/or Type CF (with define protection). Type B equipment provides a particular degree of protection against electric shock particularly regarding: m allowable leakage currents m reliability of the protective earth connection (if present). Type B equipment is, for example, suitable for intentional internal and external application to the patient, excluding direct cardiac application.
Maintenance Checks Before beginning monitoring on a patient: m Check for any mechanical damage. m Check all the external leads, plug-ins and accessories. = Check all the functions of the instrument which are needed to monitor the patient. m Ensure that the instrument is in good, working order. Do not use the HP Dravidian Telemetry Monitoring System for any monitoring procedure on a patient if you identify features which demonstrate impaired functioning of the instrument.
This is an enlarged view of the right side of the rear panel: POWER CORD CONNECTOR 2 REMOVE COVER TAB. FROM SLOT GROUNDING oR VOLTAGE SELECTOR ZLOTY V TA MA aT BR Ten woo 0.
Connectors The connectors on the rear panel of the receiver mainframe are: » Fuses: the 100/120V input voltage line is protected fuse. » Line Power Connector: this is a 3 pin connector, used to input the local line voltage. Mainframe plugin standard IEC mains inlet receptacle. w Antenna System Connector: this is a BNC coaxial connector. m Digital Data Output: these are upstream and downstream connectors that connect to the HP Dravidian monitoring network (SD.
7-8 Note 9 Warning 9 Note 9 After servicing, be certain to reconnect the secondary ground wire. Removal of the secondary grounding wire from the rear of the product voids the IEC approval. Lifting the Receiver Mainframe The HP receiver Mainframe weighs 37 lbs. (16.9 kg). When carrying the mainframe, hold it firmly from underneath. For safety reasons, it is strongly recommended that at least two people lift the mainframe. One person should not attempt it.
instructions or other information which will assist the user’s appropriate qualified technical personnel to repair those parts of the equipment which are classified by Hewlett Packard to be repairable.
Cleaning and Disinfection Warning 9 Caution $ To prevent fire, provide adequate ventilation and do not permit smoking when cleaning the transmitter or the receiver mainframe with a flammable liquid, such as alcohol, or sterilizing with ethylene oxide (ET. Disconnect line power from the receiver mainframe to prevent electrical shock and accidental turn-on. Do not use any abrasive cleaning materials on any part or component of the HP Dravidian Telemetry System.
wm Hydrogen Peroxide s Dilute sodium hypocrite (chlorine bleach), 10% solution, freshly made within the past 24 hours # Cidex® » Windex® w Lysol Make sure you rinse all cleaned surfaces with distilled water to remove any residue. Cleaning the HP Dravidian Transmitter Wiping the Outside of the Transmitter 8-2 Cleaning and Disinfection The transmitter can be cleaned by two methods: wiping or soaking. However, there are restrictions which apply to both methods.
fe Caution $ Wiping the Battery Compartment Caution $ Soaking the Transmitter Do not use Cidex®, Windex®, or Lysol inside of the battery compartment. These cleansers will damage the battery compartment. Under normal operation, the battery compartment should not become very dirty and should not require frequent cleaning.
Sterilization Hewlett-Packard makes no claims concerning the sterilization of the HP Dravidian transmitter. Cleaning ECG Patient Cables Caution $ Cleaning Caution $ The trunk cables, lead sets, and accessories that make up your HP Safety Cable System can be cleaned, disinfected, or sterilized using a variety of methods and chemicals. Please be careful not to use harsh chemicals such as Acetone. Do not immerse or soak the trunk cable or leads.
Disinfecting Caution $ Note 5 The HP cables can be disinfected when needed using the cold chemical disinfectants listed below. HP recommends that you disinfect only when necessary as determined by your hospital’s policy, to avoid long term damage to the cables, leads, etc. HP also recommends that the items being disinfected be first cleaned using the procedure described under “Cleaning”. Then, the disinfectant can be wiped onto the items using a soft cloth dampened with water, and dried with a clean cloth.
Sterilization The HP cables can be sterilized when needed using Ethylene Oxide (EtO) gas. Before sterilizing, clean the items as described under “Cleaning.” Be sure all safety precautions regarding aeration after EtO exposure are followed. HP recommends that you only sterilize these products when necessary as determined by your hospital’s policy, to help prevent long term damage to the cables leads, etc. Never autoclave or steam sterilize these products.
Caution $ M1400A/B Transmitter Sterilization Caution ¢$ Warning 9 4, Check the transducer and cable, and if you see signs of deterioration or damage, do not use for further patient monitoring. Do not autoclave the transducers. Do not use bleaches containing Sodium Hypocrite (for example, Clorox™), This generic procedure is used for sterilizing M1400A/B transmitters and associated patient cables with ethylene oxide (EtO) gas.
ETHYLENE OXIDE (29 CFR PERSONNEL EXPOSURE AND/OR ROOM AIR MUST BE MONITORED PER OSHA STANDARDS. VENT STERILIZER GAS OUTDOORS OR TO A SUITABLE, EVACUATED CONTAINER FOR REPROCESSING, DEPENDING ON STATE, PROVINCIAL, OR COUNTRY ENVIRONMENTAL REGULATIONS. DO NOT VENT STERILE INDOORS. VENT AERATOR EXHAUST ONLY TO THE OUTDOORS. These values will produce EtO residual levels in the transmitter and patient cable plastic below 250 ppm, per FDA regulations for implantable devices. Reference: U.S.A.
A Specifications and Ordering Information This appendix lists the system classification, and the environmental and electrical power specifications for the hardware components of the system. For complete specifications, see HP Dravidian Telemetry System Service Guide, part number M2600-90033. For full specifications for the wave viewer, see the HP palmtop documentation. System Classification Class 1 Equipment: M2604A Receiver Mainframe, M1401A Receiver Mainframe.
Transportation & Storage Specifications FOR ALL HARDWARE COMPONENTS OF THE HP DRAVIDIAN TELEMETRY SYSTEM EXCEPT WAVE VIEWER AND REUSABLE PULSE OPTOMETRY TRANSDUCERS: Operating Temperature Range: 32 to 113°F (0 to 45°C). Storage Temperature: -40 to +158°F (-40 to Altitude: Operating and storage up to 15,000 ft (4570 m).
Power Specifications HP M2601A Dravidian Transmitter HP M1400A/B Transmitters RF Power Output: +3 6 dBm (2 4 milliliters) nominal. Carrier Frequency Range: 460 to 470 MHz. Radio Channel Spacing: 25 kHz Defibrillator Protection: Transmitter ECG input protected against 400 joules discharge into a 50 Ohm load Batteries: 9V Alkaline, Lithium. Current Draw: 15 mA (ECG only), 56 mA, typical (ECG and Sp) Unless other indicated, specifications apply to both transmitters.
HP M2604A/M1401A Power Supply: M2604-60001, Receiver Mainframe M1403-60631 *Not applicable to M2604A. Note: See the HP Dravidian Telemetry System Service Guide for additional information on power supplies. Input Voltage: VAC delectable 10%. Frequency Range: Power Consumption: 110 VA maximum, 95 VA average, 81 W maximum, 72 W average with 8 receiver modules. Controls: Front Panel: Power On/Off; Rear Panel: Line voltage selector.
Analog Output Option (Jot) Input Voltage: USA Power Module 0950-2038: 120VAC 10%. CE Mark Power Module 0950-3221: 100-240Vac 10%. Frequency Range: Power: CE Mark Power Module 0950-3221: 33 VA maximum. Output Voltage: CE Mark Power Module 0950-3221: 24 VDC Oto 1.4 A. Output Current: CE Mark Power Module 0950-3221: 1.4 ampere DC maximum Analog Output Gain (from output of receiver module): High-level outputs: 500% + 5% Low-level outputs: 1 Inoperative Mode (IN OP Condition) Output Level: High-level output: 10.
Analog Output Card: Output (50-pin jack) Bedside Attenuation: Output (3-conductor phone jack) Holder Attenuation: Output (set of 5-button connectors) HP M1402A Receiver Frequency Range: RF Carrier 406 to 512 MHz (exact Module frequency fixed by option), VCR controlled. Channel Spacing: 25 kHz. Channel RF Bandwidth: 10 kHz. Antenna System Specifications HP M1408A Active Antenna Combiner Operating Voltage: 19 32 VDC. Current Requirements: 50 mA. Average Power Consumption: approximately 1.5 Watts.
HP M1407A Multiple Unit Power Supply Input Voltage: USA Power Module 0950-2038: 120VAC 10%. CE Mark Power Module 0950-3221: 100-240VAC 10%. Frequency Range: Power: USA Power Module 0950-2038: 36 VA maximum. CE Mark Power Module 0950-3221: 33 VA maximum. Output Voltage: USA Power Module 0950-2038: 23 VDC nominal Mark Power Module 0950-3221: 24 VDC Output Current: USA Power Module 0950-2038: 1 ampere DC. CE Mark Power Module 0950-3221: 1.4 ampere DC.
Accessories Ordering Information Accessories listed in the table below can be ordered through your Hewlett-Packard representative. See Appendix C in the HP Omnivore Component Central Monitor User’s Reference Manual for a list of Hewlett-Packard sales offices. Accessories HP Dravidian Transmitter Description HP Part Number Battery, § V Lithium (box of 10) ULBU9VLJ] Lead Set, Wire snap, 30 inch (L5H M2590A Lead Set, 3-wire grabber 30 inch (US J M25U1A Lead Set. b-wire snap, 30 inch (15.
Accessories M1400A/B Transmitter Du ascription: HP Part Number Battery, 84 volt Zine Air Box of 17 Zine Air batteries Lead Set 3 wire snap (11S) Lead Set, 3 wire grabber (U5) Lead Set, 4 wire snap (U.S) Lead Set, 4 wire grabber (U.S) Lead Set, 5 wire snap (U.S) Lead Set, 5 wire grabber (U.S) Electrode Set, disposable Pouch 1430.
Analog Output Option Warning 9 Caution $ The Analog Output Option J01 (Patient Premonitory Interface) gives the HP Dravidian Telemetry System the capability of providing ECG outputs to bedside monitors, Holder monitors, and other recording devices. The Patient Premonitory Interface (Analog Output) Option is intended for display and recording purposes only.
Analog Output Bedside Monitor Cables Note g B-2 Analog Output Option SRE TE To connect the telemetry transmitter to the bedside monitor via the optional Patient Premonitory Interface {Analog Output), you will need an analog output bedside monitor cable. The end of the cable that connects to the bedside monitor will have either a small 12-pin connector or a larger 8-pin connector (See the illustration below). The other end of the cable has a phone plug connector, and it plugs into the wall plate.
There are four different analog output bedside monitor cables. Which cable you use will depend upon whether the input connector on your bedside monitor is 8-pin or 12-pin and whether your transmitter lead set is 3-wire, 4-wire, or 5-wire. The proper cable selection is summarized in the following table. The cables can be distinguished from the 5-wire cable by the attached label (see the table on page B-5). The lead set type also determines which are the valid leads to be selected at the bedside.
Lead Placement and Selection Note $ Caution $ B-4 Analog Output Option To ensure valid waveform with the correct lead label, you must remember to use the following: » standard lead placement (shown on the telemetry transmitter case and in further detail in the section “Electrode Placement” in Chapter 2) w valid lead selection (performed at the bedside monitor) The following table summarizes recommended lead placement and selection.
Using Non-standard Lead Placement With the and 5-wire lead set, you can use non-standard lead placement, but you must still use a valid lead selection at the bedside monitor. This will give you the desired waveform, but it will result in a mislabeled lead at the bedside monitor.
Controls for Telemetry Setup To make adjustments to your patient’s ECG, you will need access to either the Patient Window Task Window on the HP Omnivore Component Central Monitor or a bedside monitor. First, go to the HP Omnivore Component Central Monitor, press and select a patient. If telemetry is not turned on, turn it on here. Then press and make adjustments. If soft key labels for telemetry controls appear, as shown below, you make all adjustments at the central monitor.
RR Functionality with paced Wave forms Caution $ In order for paced waveform to be processed correctly by bedside monitors using the analog outputs, the pace pulses must be artificially reconstructed and inserted into the analog output signals. The synthesized pace pulse is very narrow and may not be visible at the bedside display, depending on the type of monitor used.
Inoperative (IN OP) Conditions Note 9 B-8 Analog Output Option With the Patient Premonitory Interface (Analog Output) Option, the following telemetry inoperative (IN OP) conditions will appear as a LEADS OFF message at the bedside monitor. . LEADS OFF . NO SIGNAL TEL CANNOT ANALYZE REPLACE BATTERY . INTERFERENCE RECEIVER MALE NO RECEIVER TRANSMITTER MALE ECG EQUIP MALE TRANSMITTER OFF 11.
Index eet A accessories, A-7-10 analog output cables, B-1 analog output option lead placement and selection, B-3 Analog Output option, B-1 B bandwidths, 2-6 battery, 1-4 C cable 3-wire, with Dravidian transmitter, 2-14 S-wire, with Dravidian transmitter, 2-15 cables analog output, B-1 M1400A/B transmitter, 2-15 Dravidian transmitter, 2-14 Care and cleaning ECG Cables, 8-4 MI190A, 8-6 MI1900B, 8-6 Configuring the System, 6-1-8 D Prudence, 2g Index-1
Index-2 electrode application, 2-13 extended monitoring mode, 5-3 fallback mode, 5-1 with arrhythmia monitoring, 5-3 without arrhythmia monitoring, 5-1 Holder Interface option, B-1 IN OP conditions analog output option, B-8 monitoring during, 5-1 NO SIGNAL IN OP, 2-3 TEL CANNOT ANALYZE IN OP, 2-2 WEAK SIGNAL IN OP, 2-3 Installation and Patient Safety, 7-1-9 lead placement 3-wire lead set, 2-18 4-wire lead set, 2-21 S-wire lead set, 2-22 analog output option, B-3 leads M1400A/B transmitter, 2-15 Dravidian tr
ordering information, A-7-10 paced patients, 2-912 use with analog output option, B-§ power off, 1-9 radio frequency interference (RF), 2-4 receiver mainframe, 1-8 receiver module, 1-7 selecting a lead, 2-14-17 skin preparation, 2-13 specifications, 8-8 SpO; measurement physiology of, 3-1 SpO; Transducer preparation and application HP M11914A, 3-9 HP M1192A, 3-10 HP M1194A, 3-11 SpO; Transducers, 3-4 application information, 3-7 care and cleaning, 8-6 general information, 3-4 SpO; transducer selection, 3-6