User's Manual

ST/AR Arrhythmia Monitoring
6-32 ECG and Arrhythmia Monitoring
Viewing Arrhythmia Waves
To view arrhythmia beat labels:
1 Go to the Setup ECG menu.
2 Select Arrhythmia.
3 Change Annotate Arrhy from Off to On. Beat labels will be annotated
above the ECG wave and Delayed will appear beside it.
To return to the normal ECG primary lead display:
1 Select Annotate Arrhy.
2 Change to Off.
3 Exit from the Setup ECG menu.
Arrhythmia Beat Labels
Arrhythmia beat labels tell you how the monitor is classifying beats.
N = Normal
V = Ventricular Ectopic
S = Supra-ventricular Premature
P = Paced
' = Pacer spike
" = Biventricular Pacer Spike
L = Learning patient's ECG
A = Artifact (noisy episode)
? = Insufficient information to classify beats
I = Inoperative condition (e.g., LEADS OFF)
M = Pause or missed beat