User's Manual

Assigning Devices
10-4 Monitoring with other Assigned Devices
When connected the icon appears at the Information Center, Release N
or earlier.
The MX40 is assigned to the monitor. A "Tele Device Assigned" message
appears on the monitor. If the ECG wave now appears on the monitor, the
signal from the MX40 is successfully transmitting to the monitor. To
confirm that the correct MX40 has been assigned, open the ECG Setup
menu by touching the ECG waveform or HR numeric. The title of the menu
contains the equipment label of the MX40. Check that this is the correct
label.
When assigned to the monitor, the display of the MX40 appears as shown
below:
The display is primarily inactive, and there is no viewable patient data
displayed, however, battery status information is available.
If a monitor is already paired to another device, you cannot assign an MX40
to that monitor.
If the MX40 goes out-of-range or loses the short-range radio connection, it
will switch over to standard telemetry transmission to the Information
Center. In this case, the telemetry data is displayed in the Telemetry Data
Window.
If the devices are unassigned, the short-range radio connection is ended
To remove an assigned device from the MX40:
1 Press the SmartKey button.
2 Press the Add/Remove SmartKey.
3 From the Remove from menu, select the desired device and press
Confirm.
Device Assignment at the Patient Monitor
At the patient monitor, you can assign an MX40 to the patient monitor
using the Telemetry menu on the patient monitor.