User's Manual

Networked Device Synchronized Settings
10-8 Monitoring with other Assigned Devices
Networked Device Synchronized Settings
If the patient’s ECG is initially being measured with a patient monitor, and
then the patient is connected to the MX40 for monitoring, the Information
Center will use the patient monitor settings for the MX40. When the initial
ECG source is the MX40, and then the patient is connected to the monitor,
the Information Center uses its Telemetry Setup settings. The following
settings will be synchronized:
Heart Rate
HR/Pulse Alarm On/Off, Heart Rate High/Low Limit
ECG
Primary Lead, Secondary Lead, Va Lead, Vb Lead
Arrhythmia
On/Unlocked, On/Locked, Off/Unlocked, Off/Locked,
Off/Not Available, Analysis Mode, Asystole
Threshold, Pause Threshold, VTach HR, VTach
Run, PVCs/min, Vent. Rhythm, SVT HR, SVT Run,
PVCs/min On/Off, Pacer not Capture On/Off, Pacer
not Pace On/Off, Non-sustain On/Off, Vent. Rhythm
On/Off, Run PVCs On/Off, Pair PVCs On/Off,
Missed Beat On/Off, Pause On/Off, R-on-T On/Off,
Vent. Bigeminy On/Off, Vent. Trigeminy On/Off,
Multiform PVCs On/Off, Irregular HR On/Off, SVT
On/Off, Afib On/Off, End Afib Time, Afib Remind
Time
ST
ST Analysis On/Off, ST Alarm On/Off, ISO point, J
point, ST point, ST Priority List, Single ST Alarm
Limit, Multi ST Alarm Limit, ST Auto/Manual
STE
STE Analysis On/Off, STE Alarm On/Off
QT
QT Analysis On/Off, QTc High On/Off, QTc High
Alarm Limit, dQTc High On/Off, dQTc High Alarm
Limit, QT Lead, QTc Correction Formula, QT
Baseline
SpO
2
T
SpO
2
Alarms On/Off, SpO
2
Alarm Limits, NBP Alarm
Suppression On/Off, Pulse (SpO
2
) On/Off, Desat
Limit, SpO
2
Averaging Time