Patient Configurable Settings in Telemetry Setup (IIC)
Monitoring with the MX40 at the Information Center 11-11
Patient-Configurable Settings in Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Telemetry/
Multi-Function
Button
Determines the
Information Center
response when the
Multi-Function Button is
pressed.
Nurse Call - generate nurse call
alarm that can be retrieved from
Alarm Review for later use.
Record - generate a recording strip
Nurse Call and Record - generate
nurse call alarm and recording strip
None
Nurse Call
Fixed Pacer
Amplitude
Sets the appearance of
the pacer spikes to a
fixed size as they
appear in the patient
window.
enable
disable
Note this does not affect the
appearance of the pacer spikes on
the MX40.
disable
SpO
2
Enabled
Enable/disable the
SpO
2
measurement at
the Information Center
and the MX40.
enable
disable
enable
SpO
2
Mode
Determine the MX40
SpO
2
measurement
behavior.
Note Pulse Rate
and Pleth Wave are not
available in Spot Check
(Manual) mode.
Spot Check (Manual)- Provides
manual measurements so the
clinician can check as needed.
Measurement can be initiated at the
MX40 from the SmartKeys menu or
the SpO
2
Setup menu and at the
Information Center by selecting the
Spot Check SpO
2
icon in the Patient
Window.
Continuous - Sends an SpO
2
parameter value to the Information
Center every second. If selected,
Pulse Rate and Pleth Wave may also
be sent.
Spot Check
(Manual)