User's Manual

Unit Configurable Settings (IIC)
11-16 Monitoring with the MX40 at the Information Center
Unit Settings - Telemetry Setup
Control
Function
Settings
Factory
Default
SpO
2
Mode
Determine the MX40
SpO
2
measurement
behavior.
Note Pulse Rate and
Pleth Wave are not
available in Spot Check
(Manual) mode.
Spot Check (Manual)-
Provides manual
measurements so the clinician
can check as needed.
Measurement can be initiated
from the SmartKeys menu, the
SpO
2
Setup menu or by
selecting the Spot Check
SpO
2
icon in the Patient
Window.
Continuous - Sends an SpO
2
parameter value to the
Information Center every
second. If selected, Pulse
Rate and Pleth Wave may
also be sent.
Spot Check
(Manual)
Suppress SpO
2
Inops with NBP
Enable/disable the SpO
2
algorithm to detect NBP
running and suppress
sending technical alarms
from the MX40 for 60
seconds.
Warning
If you measure SpO
2
on a
limb that has an inflated
NBP cuff, a non-pulsatile
SpO
2
technical alarm can
occur. If the monitor is
configured to suppress
this alarm, there can be a
delay of up to 60 seconds
in indicating critical
patient status, such as
sudden hypoxia.
enable
disable
enable