User's Manual

Unit Configurable Settings (IIC)
Monitoring with the MX40 at the Information Center 11-17
Unit Settings - Telemetry Setup
Control
Function
Settings
Factory
Default
Pleth Wave
Enable/disable the
transmission of the Pleth
wave and its subsequent
display to the Information
Center. For Continuous
mode only.
enable
disable
Note When enabled,
during 6-lead monitoring, the
Pleth wave will replace the Vb
wave in the Patient Window.
disable (Pleth
wave is not
displayed.)
Pulse
Enable/disable the
transmission of the Pulse
rate and its subsequent
display to the Information
Center. For Continuous
mode only.
enable
disable
disable (Pulse
rate is not
displayed.)
SpO
2
Alarm
Turn SpO
2
alarms on/off
at the Information Center.
enable (on)
disable (off)
enable
SpO
2
Limits
High
Increment/decrement
SpO
2
high alarm limit by 1
(in %).
Limit maximum is 100. Limit
minimum is 51 (adult) or 31
(pediatric). High and low limit
must be at least 1% apart.
100 (adult,
pediatric)
SpO
2
Limits
Low
Increment/decrement
SpO
2
low alarm limit by 1
(in %).
Limit maximum is 99. Limit
minimum is 50 (adult) or 30
(pediatric). High and low limit
must be at least 1% apart.
90 (adult,
pediatric)