Global Settings (IIC iX)
Monitoring with the MX40 at the Information Center 11-21
Global Settings (IIC iX)
The following tables identify the settings (and available choices) through
the Alarm Management, Telemetry Setup, and ECG Management menus
from the Global Settings application page at the IntelliVue Information
Center iX. For more information, including information on Telemetry
Profiles, see the IntelliVue Informaiton Center iX Clinical Configuration Guide,
Part Number 453564292621.
Alarm Management
Control
Function
Setting Choices
Factory
Default
Alarms Off Priority
Sets the types of alarms that can
be paused for a configured
amount of time from the IIC iX..
Red and Yellow
Yellow Only (red alarms must be
paused at the device)
Not Allowed
Yellow Only
Alarms Off
Sets the amount of time all alarms
will be paused if selected.
1, 2, or 3 minutes.
2 min
Audible Latching
Sets which alarms must be
silenced even if the condition no
longer exists (applies to
non-arrhythmia alarms only).
Red and Yellow
Red Only
Red&Yellow
Alarm Reminder
Sets the appearance and
behavior of how an alarm will
remind if the alarm condition
consists (applies to
non-arrhythmia alarms only).
On
Realarm
Off
On
Inop Reminder
Sets the appearance and
behavior of how an Inop will
remind if the Inop condition
consists
On
Realarm
Off
On
Reminder Time
Sets the time period for
Reminders.
1, 2, or 3 minutes
3 min
No Data Inop
Sets the severity of the NO DATA
FROM MONITOR inop.
Hard (always for MX40)
Soft
Hard
ECG Leads Off
Sets the severity for the ECG
LEADS OFF Inop
Cyan
Yellow
Red
Cyan
Replace Battery
Sets the severity for the
REPLACE BATTERY Inop.
Cyan
Yellow
Red
Cyan
Some ECG Al Inop
Allows for notification whenever
the On/Off settings for
ECG/Arrhythmia alarms differ
from the current Profile.
On
Off
On