User's Manual

Global Settings (IIC iX)
11-22 Monitoring with the MX40 at the Information Center
Control
Function
Setting Choices
Factory
Default
HR Alarms
Sets heart rate limit alarm type.
Short Yellow
Yellow
Short Yellow
Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Telemetry Button
Determines the action when the
Telemetry Button is pressed.
Nurse Call - generates nurse call
alarm that is marked as an event
in Alarm Review.
Record - generates an automatic
recording.
Call & Record - Creates both the
nurse call alarm and automatically
starts a recording.
Off
Nurse Call
Standby Duration
Sets the standby duration time
when Standby is selected at the
IIC iX.
Standby Duration is always
Infinite when selected at the
MX40.
10 minutes
20 minutes
30 minutes
1 hour
2 hours
3 hours
4 hours
Infinite
Infinite
Battery Gauge
Selects whether a battery gauge
is displayed on the IIC iX for
devices operatimg on battery
power.
On
Off
On
Remote Pause
Enables all alarms to be paused
at the device for the same
configured amount of time that
they are paused at the
Information Center.
On
Off
Off
RF Auto Shutoff
Selects whether the device will
shut off after ECG LEADS OFF
for longer than 10 minutes and
the SpO
2
sensor cable is not
connected for longer than 10
minutes.
On
Off
On
SRR Use Model
Selects whether the device will
assing to patient monitor or
cableless measurement device.
Look for Sensor
Look for Monitor
Look for
Monitor
SRR Fast Transition
Reduce the delay time when fast
switching between the MX40 and
the patient monitor
On
Off
On