User's Manual

Charging Lithium-ion Rechargeable Batteries
Maintenance 14-9
Status
LED
no battery on charger slot
off
battery put on charger slot
flashing yellow
battery not properly recognized, error
cyan
battery recognized, battery charging
yellow
battery recognized, battery full (>90%)
green
The AC Power / Error LED is
green when the charging station is connected to AC power
cyan during startup or to indicate a general charging station error
Note Wiping of battery contacts with an alcohol solution after cleaning is
recommended.
Battery Status on the Charging Station Display
The IntelliVue CL Charging Station display provides a quick overview of
all the connected devices and their battery status. The screen is arranged in
the same layout as the charger slots.
Battery Lifetime Management
The lifetime of a li-ion battery depends on the frequency and duration of
use. When properly cared for, the useful life is approximately 4 years or 500
complete charge-discharge cycles, whichever comes first. In addition,
experience indicates that the incidence of failure may increase with battery
service life due to the accumulated stresses of daily use. We therefore
strongly recommend that li-ion batteries be replaced after 2 years or 500
complete charge-discharge cycles.
The age of a li-ion battery begins at the date of manufacture. The date of
manufacture is listed on the side of the battery.