User's Manual

Electromagnetic Compatibility
Safety Standards & Specifications 15-11
Recommended Separation Distance
Warning
The MX40, equipped with a wireless network interface, intentionally
receives RF electromagnetic energy for the purpose of its operation.
Therefore, other equipment may cause interference, even if that other
equipment complies with CISPR emission requirements.
In the following table, P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Portable and mobile RF communications equipment should be used no
closer to any part of the MX40, including cables, than the recommended
separation distance calculated from the equation appropriate for the
frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of equipment marked with this
symbol:
Immunity Test
IEC 60601-1-2 Test
Level
Compliance Level
Electromagnetic
Environment Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 VRMS
Recommended separation
distance:
d = 1.2√P
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Recommended separation
distance:
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P