User's Manual

Battery Specifications
15-16 Safety Standards & Specifications
Monitor Mode
Non-networked
(Display On)
Battery Life
(1.4GHz
p/n 865350)
Battery Life
(2.4GHz
p/n 865351)
Battery Life
(WLAN
p/n 865352)
ECG/SpO
2
Manual
In this mode battery life is dependent on the usage rate and will
range between the ECG Only battery life and the ECG/SpO
2
Continuous battery life.
Note Use of the short-range radio can reduce battery life by 25%.
Note The battery capacity of re-chargeable batteries degrades over time
and number of recharge cycles. Toward the end of its useful life, the battery
capacity may be reduced by 25-30%. If this reduced battery life is
unacceptable based on your use model, Philips recommends replacing the
rechargeable battery sooner.
Nominal Current
Operating Mode
Nominal Current
(p/n 865350)
Nominal Current
(p/n 865351)
Nominal Current
(p/n 865352)
ECG Only (Display
inactive)
67 mA @ 3.6V
67 mA @ 3.6V
70 mA @ 3.6V
ECG/SpO
2
Continuous
(Display inactive)
136 mA @ 3.6V
136 mA @ 3.6V
140 mA @ 3.6V