User's Manual

MX40 2.4 GHz Smart-Hopping Radio
15-20 Safety Standards & Specifications
MX40 2.4 GHz Smart-Hopping Radio
Parameter
Specification
Frequency Range
ISM Band: 2400 - 2483.5 MHz
Channel Assignment
48 radio channels assigned from 2401.056 MHz
- 2482.272 MHz
Channel Spacing: 1.728 MHz
RF Output Power
FCC: Channels 0-46 -17 dBm +/- 1 dB (40 mW
to 63 mW, nominal 50 mW), into antenna load.
Channel 47 only - 15 dBm +/- 1 dB.
ETSI: 12 dBm +/- 1 dB (13 mW to 20 mW,
nominal 16 mW), into antenna load
ARIB: 13.5 dBm +/- 1 dB (18 mW to 28 mW,
nominal 22 mW), into antenna load
Radio Frequency Accuracy
during normal operation
<+ 60 /- 100 KHz relative to channel frequency,
includes temperature compensation and aging
effects
Modulation Type
GFSK, Gaussian Frequency Shift keying
(1M40Q7D)
Modulation Bandwidth
Typically 1.4 MHz (20 dB Bandwidth)
Typically 980 KHz (6 dB Bandwidth)
Out of Band Spurious
Emission Levels
Meets ETSI, RS210, FCC, ARIB standards
2.4 GHz ISM
FCC and Industry Canada Radio Compliance: This device complies with
Part 15 of the FCC Rules and RSS-210 of Industry Canada. Operation is
subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation. Any
changes or modifications to this equipment not expressly approved by
Philips Medical Systems may cause harmful radio frequency interference
and void your authority to operate this equipment.