User's Manual

Measurement Specifications
Safety Standards & Specifications 15-27
Parameter
Specification
Pacer Rejection
Performance
(Pace pulses with no
tails).
Positive pacers
1
Amplitude Width
+2 to +700 mV 0.1, 0.2, 0.5 and 1.0 ms
+2 to +500 mV 1.5 ms
+2 to +400 mV 2 ms
Negative pacers
1
Amplitude Width
-2 to -700 mV 0.1, 0.2, 0.5 and 1.0 ms
-2 to -500 mV 1.5 ms
-2 to -400 mV 2 ms
1
Philips does not claim, verify, or validate support
for all available pacemakers.
EMC Performance Limits,
radiated immunity
Meets Essential Performance.
ECG Patient Cable
Disconnection Safety
All ECG connections are patient safe within 750
msec of patient cable removal, with patient leakage
current <10  Exception: Leadset detection pins
are protected mechanically to prevent patient
contact.
ECG Performance Disclosure/Specifications
Characteristic
Performance Disclosure/Specification (in
italics)
Heart Rate Averaging Method
Two different methods are used:
Normally, heart rate is computed by averaging the 12
most recent RR intervals.
If each of 3 consecutive RR intervals are greater than
1200 milliseconds (i.e. rate less than 50 b/min) for adult
and pediatric patients, then the 4 most recent RR
intervals are averaged to compute the HR.
Heart Rate Meter Accuracy and Response
to Irregular Rhythm
Provides correct heart rates (80, 60, 120, 90 b/min) using
test waveforms as indicated in ANSI/AAMI EC13
Sec. 4. 1. 2. 1 (e).
Response Time of Heart Rate Meter to
Change in Heart Rate
For a rate increase, the average time to reach the specified
heart rate using test waveforms as indicated in ANSI/AAMI
EC13 Sec. 4. 1. 2. 1 (f) is 10 seconds. For a rate drop, the
average time is 7 seconds.