User's Manual

Measurement Specifications
Safety Standards & Specifications 15-29
Characteristic
Performance Disclosure/Specification (in
italics)
Frequency Response: Triangular
Meets the ANSI/AAMI EC13 Section 4.2.9.8(b) standard: 0
to 25% reduction.
Impulse Response:
(for waves marked with ST bandwidth)
Meets the ANSI/AAMI EC13 Section 4.2.9.8(c) standard:
displacement maximum = 0.1 mV; slope
maximum = 0.30 mV/s.
Pacemaker Pulse Display Capability
Meets the ANSI/AAMI EC13 Section 4.2.9. 12 standard:
minimum = 0.2 mV RTI.
Tall T-Wave Rejection Capability
Meets AAMI standard:
0.5 40 BW: HR of 80bpm at all T-wave amplitudes
0.05 40 BW: HR of 80bpm at all T-wave amplitudes
Respiration
Parameter
Specification
Leads Used for Measurement
RA, LL (standard) or I, A (EASI)
Range
Adult/Pedi: 0 to 120 rpm
Bandwidth
0.3Hz to 2.5Hz (-6dB)
Noise
Less than 25 mOhm (rms) referred to the
input
Calibration Signal
Signal: 1 Ohm p-p; Accuracy: +/- 20%
Respiration Rate Resolution
1 rpm
Respiration Accuracy
+/- 1 rpm for 0-120 rmp
Auxiliary Current, Respiration Excitation
Signal
< 470 uA rms @48KHz, sinusoidal waveform
Respiration Alarm
Alarm
Range
Delay
High
Adult/Pediatric: 10 to 100 rpm
≤ 15 seconds
Low
Adult/Pediatric: 0 to 95 rpm
for limits from 0 to 20 rpm: max. 4 seconds
for limits above 20 rpm: max. 15 seconds
Apnea Alarm
10 to 40 seconds
Incremental delay 5 seconds max.