Physiologic Alarms
Alarms 5-11
Alarm Text
Priority
Condition
Source
*** DESAT
Red
Very Low SpO
2
Saturation.
SpO
2
value below Desaturation limit
Note Desat limit is set 10 points below
low limit with IIC Release N. It is
configurable with IIC iX.
SpO
2
***EXTREME TACHY
Red
Extreme Tachycardia.
Heart Rate (yyy) greater than Extreme
Tachy limit
ST/AR Basic
& Enhanced
Arrhythmia
*** VENT FIB/TACH
Red
Ventricular Fibrillation.
Fibrillatory wave (sinusoidal wave
between 2-10 Hz) for 4 consecutive
seconds
ST/AR Basic
& Enhanced
Arrhythmia
*** V-TACH
Red
Ventricular Tachycardia.
Consecutive PVCs greater than or equal
to V-Tach Run limit and Heart Rate
greater than V-Tach limit (xxx)
ST/AR Basic
& Enhanced
Arrhythmia
** dQTc
Yellow
dQTC value has exceeded the dQTC high
limit for > 5 minutes for 2 consecutive time
periods.
ECG/QT
** MULTI ST
Yellow
Two contiguous ST leads have values
exceeding elevation or depression alarm
limits for > 60 sec.
ECG/ST
** NBP High
Yellow
High limit has been exceeded for high
pressure limit.
NBP
** NBP Low
Yellow
Low limit has been exceeded for low
pressure limit.
NBP
** QTc High
Yellow
QTc value has exceeded the QTc high
limit for > 5 minute for2 consecutive time
periods.
ECG/QT
** RR HIGH
Note Only available with
IIC iX.
Yellow
The respiration rate has exceeded the
high alarm limit.
RESP
** RR LOW
Note Only available with
IIC iX.
Yellow
The respiration rate has dropped below
the low alarm limit.
RESP
**SpO
2
T HIGH
Yellow
High SpO
2
.
SpO
2
value (yyy) greater than high SpO
2
limit (xxx)
SpO
2
** SpO
2
T LOW
Yellow
Low SpO
2
.
SpO
2
value (yyy) less than low SpO
2
limit
(xxx).
SpO
2
** ST-LEAD HIGH
Yellow
ST segment is elevated.
ECG/ST
** ST-LEAD LOW
Yellow
ST segment is depressed.
ECG/ST