User's Manual

Technical Alarms (INOPs)
Alarms 5-15
Alarm Text
Priority
Condition
What to do
CHECK PAIRING
Source - MX40
Yellow
Technical
Alarm
There is a problem with
device pairing.
When the MX40 is
wirelessly paired with an
X2 patient monitor (no
label) docked with a
larger networked MP
series monitor, and the
network connection is
lost.
Check that the bedside
monitor or cableless
measurement device is
correctly paired.
Select the correct device to be
paired.
cl NBP Batt Low
Source - Cableless
Measurement Device
Hard
CL NBP Pod weak battery
condition.
Charge CL NBP Pod.
cl NBP Batt Empty
Source - Cableless
Measurement Device
Severe
CL NBP Pod empty battery
condition. Monitoring is not
possible.
Replace CL NBP Pod.
Recharge depleted CL NBP
Pod.
cl NBP DISCONNECT
Source - Cableless
Measurement Device
Hard
CL NBP Pod is not
connected with the MX40.
Resolve interference
condition.
Reduce range between CL
NBP Pod and MX40.
cl SpO
2
Batt Low
Source - Cableless
Measurement Device
Hard
CL SpO
2
Pod weak battery
condition.
Charge CL SpO
2
Pod.
cl SpO
2
Batt Empty
Source - Cableless
Measurement Device
Severe
CL SpO
2
Pod empty
battery condition.
Monitoring is not possible.
Replace CL SpO
2
Pod.
Recharge depleted SpO
2
Pod.
cl SpO
2
DISCONNECT
Source - Cableless
Measurement Device
Hard
CL SpO
2
Pod is not
connected with the MX40.
Resolve interference
condition.
Reduce range between CL
SpO
2
Pod and MX40.
!!!/!! CUFF NOT
DEFLATED
Source - Cableless
Measurement Device
Severe
Cuff pressure has
exceeded the specified
safety limit.
Remove cuff and tubing and
expel air.
!!!/!! CUFF OVERPRESS
Source - Cableless
Measurement Device
Severe
Cuff pressure has
increased above
overpressure safety limits.
Remove cuff and tubing and
expel air.