User's Manual

Technical Alarms (INOPs)
5-16 Alarms
Alarm Text
Priority
Condition
What to do
ECG/ARRH ALARM(S)
OFF
Source - MX40
Soft
ECG is turned off.
Turn on ECG.
ECG LEADS OFF
Note This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Source - MX40
Red or
Yellow or
Hard
Technical
Alarm
Multiple leads are off.
Re-attach ECG leads to patient..
<electrode> LEAD OFF
Source - MX40
Hard
Single lead is off.
If primary lead is MCL,
lead will be identified as
V/C in INOP text.
Re-attach ECG leads to patient.
LEADSET UNPLUGGED
Source - MX40
Hard
Patient cable has been
unplugged from the
MX40.
Incompatible leadset
attached to patient cable.
Re-attach the patient cable.
Replace the leadset.
LEADSET LIFE
Soft
The single-patient use
leadset has exceeded its
limit of 25 cycles.
Replace with new leadset.
LOCAL AUDIO OFF
Source - MX40
Note This is normal
operation in Telemetry
Mode.
Soft
There is no alarm audio
notification when operating
in Telemetry Mode.
Change to Monitor Mode.
NBP INTERRUPTED
Source - Cableless
Measurement Device
Hard
The preset maximum time
for the total measurement
has been exceeded.
Reduce patient movement and
avoid interaction with the cuff
and tubing.
NBP MEASURE FAILED
Source - Cableless
Measurement Device
Hard
Measurement values
cannot be derived.
Attach cuff to new location on
patient.
Replace cuff.