User's Manual

Technical Alarms (INOPs)
5-20 Alarms
Alarm Text
Priority
Condition
What to do
TELE MALFUNCTION
Source - MX40
Hard
MX40 malfunction or
self-test failure.
Contact Service to replace the
MX40.
TELE REMOVE BATT
Source - MX40
Note For Information
Center Release L or M, this
INOP will appear as
"REPLACE BATTERY T".
Hard,
Latched
The temperature of the
lithium-ion battery is >60
o
C and the battery must be
removed.
Replace the lithium-ion
battery.
Dispose of old battery
properly.
TELE SERVICE BATT
Source - MX40
Hard
The lithium-ion battery has
exceeded the maximum
charge cycle limit and
reached the end of its
useful life.
Replace the lithium-ion
battery.
Dispose of old battery
properly.
TELE WEAK SIGNAL
Source - MX40
Soft
Patient is at outer range
of the radio coverage
area.
The MX40 is receiving a
weak signal with high
data loss from the AP.
Condition exists for
multiple devices in a
specific area
Return patient to the coverage
area.
If patient is in close proximity
to AP, replace the MX40.
Contact service.
The AP covering the specific
area is suspect. Contact
Service
TRANSMITTER OFF
Source - MX40
Hard
RF Auto Shutoff after 10
minutes of all leads off and
no SpO
2
sensor
connected.
Reattach ECG leads to
patient.
Reattach SpO
2
sensor.
UNSUPPORTED LAN
Source - MX40
Hard
The MX40 (WLAN) is
connected to the Access
Point, but cannot obtain an
IP address.
Correct the IP address issue.
WIFI OUT OF RANGE
Source - MX40
Hard
The MX40 (WLAN) is out
of range of an access
point.
Correct the RF coverage issue.