User's Manual

Choosing EASI or Standard Lead Placement
ECG and Arrhythmia Monitoring 6-13
Choosing EASI or Standard Lead Placement
Choose either standard lead placement or EASI lead placement:
1 In the Setup ECG menu, select Lead Placement to toggle between
Standard or EASI.
2 Select Standard or EASI.
Note When changing lead placement, the patient cable must be attached
to the MX40.
EASI is shown beside the 1 mV calibration bar on the ECG wave on the
display, and EASI is marked on any recorder strips and printouts.
See the sections on Lead Placement for electrode placement diagrams.