User's Manual

9-8 Monitoring with the MX40 at the Information Center
Patient-Configurable Settings in Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Telemetry/
Multi-Function
Button
Determines the
Information Center
response when the
Multi-Function Button is
pressed.
Nurse Call - generate nurse call
alarm that can be retrieved from
Alarm Review for later use.
Record - generate a recording
strip
Nurse Call and Record - generate
nurse call alarm and recording
strip
None
Nurse Call
Fixed Pacer
Amplitude
Sets the appearance of
the pacer spikes to a
fixed size as they
appear in the patient
window.
enable
disable
Note this does not affect the
appearance of the pacer spikes
on the MX40.
disable
SpO
2
Enabled
Enable/disable the
SpO
2
measurement at
the Information Center
and the MX40.
enable
disable
enable
SpO
2
Mode
Determine the MX40
SpO
2
measurement
behavior.
Note Pulse Rate
and Pleth Wave are not
available in Spot Check
(Manual) mode.
Spot Check (Manual)- Provides
manual measurements so the
clinician can check as needed.
Measurement can be initiated at
the MX40 from the SmartKeys
menu or the SpO
2
Setup menu
and at the Information Center by
selecting the Spot Check SpO
2
icon in the Patient Window.
Continuous - Sends an SpO
2
parameter value to the
Information Center every second.
If selected, Pulse Rate and Pleth
Wave may also be sent.
Spot Check
(Manual)