User's Manual Addendum

Addendum to the Instructions for Use
3
1English Addendum to the Instructions for Use
This addendum provides additional or corrected information for the following IntelliVue Instructions
for Use (software revision G.01.xx):
MP20-MP90 Patient Monitor (M8000-9001K)
MP5 Patient Monitor (M8105-9001C)
MP2 Patient Monitor (M8102-9001B)
X2 Multi-Measurement Module (M3002-9001B)
If your Instructions for Use has a different part number, you may discard this addendum, otherwise
please store it with your monitor documentation.
CO
2
Alarm Specifications
For Microstream CO
2
in the MP5 and for the M3015A, the CO
2
alarm specifications are as follows:
IntelliVue Telemetry Transceiver User Information
For complete user information, including accessory listings and specifications, for the M4841A/
M4851A TRx/TRx+ IntelliVue Telemetry Transceiver, please refer to the Instructions for Use
provided with the device.
MP2/X2 Including the Wireless IntelliVue Adapter
Radiofrequency Radiation Exposure Information
For body worn operation, this equipment has been tested and meets the FCC RF exposure guidelines
when used with the accessories supplied or those approved for use with this product. Use of other
accessories may not ensure compliance with FCC RF exposure guidelines.
CO
2
Alarm
Specifications
Range Adjustment Delay
etCO
2
High 20 to 95 mmHg (2 to 13 kPa) 1 mmHg (0.1 kPa) less than 21 seconds.
etCO
2
Low 10 to 90 mmHg (1 to 12 kPa)
imCO
2
High 2 to 20 mmHg
(0.3 to 3.0 kPa)
steps of 1 mmHg
(0.1 kPa)
less than 21 seconds.
awRR High Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
under 20 rpm: 1 rpm
steps
over 20 rpm:5 rpm
steps
less than 21 seconds.
awRR Low Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
settings <20 rpm: less than 8 seconds
>20 rpm: less than 21 seconds
Apnea delay 10 to 40 seconds 5 second steps set apnea delay time + 8 seconds (with
2 m sample lines) or 11 seconds (with
4 m sample lines).