053621 EN, DE DSF 4/16/08 EverFlo™ EverFlo™ Q
English EverFlo / EverFlo Q User Manual EverFlo / EverFlo Q User Manual Table of Contents Symbol Key ....................................................................................................... 2 Abbreviations ................................................................................................... 2 Chapter 1: Introduction ................................................................ 3 Intended Use ..................................................................................
EverFlo / EverFlo Q User Manual Symbol Key Follow Instructions for Use AC Power Type BF Applied Part REF Model Number Class II equipment SN Serial Number No smoking On (Power) No oil or grease Off (Power) Do not disassemble IPX1 Drip proof equipment General Alarm Action Required, Check System Notification European Declaration of Conformity Compliant with the Waste Electrical and Electronic Equipment/Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipm
EverFlo / EverFlo Q User Manual English Chapter 1: Introduction Your health care professional has determined that supplemental oxygen is of benefit to you and has prescribed an oxygen concentrator set at a specific flow setting to meet your needs. DO NOT change the flow settings unless your health care professional tells you to do so. Please read and understand this entire manual before using the device.
EverFlo / EverFlo Q User Manual Warnings and Cautions Warnings A warning represents the possibility of harm to the operator or patient. • For proper operation, your concentrator requires unobstructed ventilation. The ventilation ports are located at the rear base of the device and at the side air inlet filter. Keep the device at least 15 to 30 cm away from walls, furniture, and especially curtains that could impede adequate airflow to the device.
EverFlo / EverFlo Q User Manual Chapter 2: Operating Instructions English Oxygen Outlet Port Warning: Do not use extension cords or electrical adapters. 1. Select a location that allows the concentrator to draw in room air without being restricted. Make sure that the device is at least 15 to 30 cm away from walls, furniture, and especially curtains that could impede adequate airflow to the device. Do not place the device near any heat source. 2.
EverFlo / EverFlo Q User Manual 4. Press the power switch to the On [I] position. Initially, all the LEDs will illuminate and the audible alert will beep for a few seconds. After that time, only the green LED should remain lit. You can begin breathing from the device immediately even though it typically takes 10 minutes to reach oxygen purity specifications. 5.
EverFlo / EverFlo Q User Manual English Chapter 3: Cleaning & Maintenance Warning: It is important to unplug the device before you perform any cleaning. Caution: Excess moisture may impair the proper operation of the device. Cleaning Periodically, use a damp cloth to wipe down the exterior case of the EverFlo / EverFlo Q device. If you use medical disinfectants, be sure to follow the manufacturer's instructions.
EverFlo / EverFlo Q User Manual Chapter 4: Alarms and Troubleshooting Alarm and Indicators The device has an audible alarm and three LED indicators, as shown below. NORMAL Green Power On Indicator Yellow Check System Indicator Red Alarm Indicator 8 Audible Alarm / Colored LED Possible Cause Your Action All 3 LEDs illuminate continuously and the Audible Alarm is sounding continuously. The device has detected a system malfunction.
EverFlo / EverFlo Q User Manual Problem Why it Happened What to Do Yellow LED is blinking. The Red LED is off and the Audible Alarm is beeping periodically. The device has detected a high oxygen flow condition. Turn the flow rate down to your prescribed level. Wait at least 2 minutes. If the condition persists turn the unit off, connect to a back up oxygen source, and call your home care provider. Green LED illuminates continuously. The other LEDs are off and the Audible Alarm is silent.
EverFlo / EverFlo Q User Manual Chapter 5: Specifications Environmental Operating Transport & Storage 13 to 32°C -34 to 71°C 15 to 95%, noncondensing 15 to 95%, noncondensing 0 to 2286 m N/A Temperature Relative Humidity Altitude Physical Dimensions 58 cm x 38 cm x 24 cm Weight 14 to 15 kg Standards Compliance This device is designed to conform to the following standards: - IEC 60601-1 Medical Electrical Equipment, Part 1: General Requirement for Safety - IEC 60601-1-2 2nd edition, Medical Electr
EverFlo / EverFlo Q User Manual English Oxygen Oxygen Concentration* (All Models except 90-96% from 0.5 to 5 LPM 1020007, 1020008) Models 1020007, 1020008 87-96% from 0.5 to 5 LPM * Device operation above or outside of the voltage, LPM, temperature, humidity and/or altitude values specified may decrease oxygen concentration levels.
EverFlo / EverFlo Q User Manual Appendix A: EMC Information GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS: This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT – GUIDANCE RF emissions CISPR 11 Group 1 The device uses RF energy only for its internal function.
EverFlo / EverFlo Q User Manual GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY: This device is intended for use in the IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL English electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
EverFlo / EverFlo Q User Manual Limited Warranty Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of three (3) years from the date of sale by Respironics, Inc. to the dealer.