Date:
15 Jan 2014
Field communication:
16-700-092
System One 50 and 60 Series Blower
Assembly Update
Page 1 of 4
FRM 2106
Revision: 05
From:
Erica Hursh
Engineering Specialist
To:
Product(s):
The following System One 50 and 60
Series (2.XX and 3.XX only) models are
affected:
PR System One REMstar (except 15xP
devices remaining at firmware v2.53)
PR System One REMstar Plus
PR System One REMstar Pro w/ C-Flex+
PR System One REMstar Auto w/ A-Flex
PR System One BiPAP Pro
PR System One BiPAP Auto
Action
required:
Subject:
System One 50 and 60 Series Blower Assembly Update
Purpose / Scope:
This Field Communication provides details regarding a Blower Assembly update, along with instructions on connecting the
updated Blower Assembly to the PCA.
Detail / Description:
Blower Assembly (PN 1080758) has been updated to remove the thermistor (2-pin connector with 2 wires). This update was
made due to the thermistor being deemed an unnecessary component. There are no adverse effects to the devices, as this
update does not affect device function or performance. Refer to the next page for illustrations on the updated Blower
Assembly.
*Note: The part number to this Blower Assembly has not changed the update was only made to the component
itself.
**Note: 150P and 151P devices remaining at firmware v2.53, as per Field Communications 16-700-001 and 16-700-021,
will no longer use Blower Assembly (PN 1080758). When replacing the Blower Assembly on these devices, use PN
1083248. Refer to Field Communication 16-700-099 for details on this part number change.