HeartStart FRx Trainer INSTRUCTIONS FOR USE 861306 Edition 5
INTENTIONALLY BLANK P H I L I P S M E D I C AL S Y S TE M S
HeartStart FRx Trainer 861306 INSTRUCTIONS FOR USE Edition 5
About this edition Device manufacturer The information in this guide applies to the HeartStart FRx Trainer 861306. This information is subject to change. Please contact Philips at www.philips.com/productdocs or contact your local Philips representative for information on revisions. Philips Medical Systems, Seattle, WA, 98121-1825, USA Edition history Trademarks Little Anne CPR Manikin and Resusci Anne CPR Manikin are trademarks or registered trademarks of Laerdal Medical.
HEARTSTART FRX TRAINER 861306 PHILIPS MEDIC AL SYSTEMS CONTENTS Intended use ........................................................................................... 2 Features ................................................................................................... 2 Overview ................................................................................................ HeartStart Trainer Kit Contents ................................................ Optional Accessories ................
INTENDED USE The HeartStart FRx Trainer is designed to prepare emergency responders to use the FRx automated external defibrillator (AED). The HeartStart FRx Trainer cannot be used to deliver defibrillation treatment. The FRx Trainer provides a variety of simulations, or training scripts, to help responders become familiar with the FRx defibrillator and allow them to practice the basic skills necessary to use the defibrillator in an emergency.
• • The HeartStart FRx Trainer has eight training scenarios that simulate realistic sudden cardiac arrest episodes. See “Standard Training Scenarios” for scenario descriptions. These scenarios are compatible with training programs developed by nationally recognized responder programs. The HeartStart FRx Trainer is powered by four standard AA alkaline batteries. WARNING: Do not store the Trainer with your defibrillator, or where it could be confused with a defibrillator during an emergency.
A TRAINING PADS CONNECTOR PORT. The Training Pads II connector is inserted in this port. B INFANT/CHILD KEY SLOT. The Infant/Child Key, available separately, is inserted into this slot to simulate a defibrillation scenario for an infant or child under 8 years old or 55 pounds. C ON LIGHT. The FRx Trainer's green ON light is on solid when the Trainer is on, and off otherwise. (In the FRx Defibrillator, the green Ready light blinks when the defibrillator is in standby mode.) D ON/OFF BUTTON.
K BATTERY COMPARTMENT DOOR. This door is lifted off in order to install or replace batteries. The Trainer is powered by four disposable AA alkaline cells inserted in the recess on the back of the unit. (The FRx Defibrillator is powered by a custom battery pack.) L CONFIGURATION SWITCHES. Located inside the battery compartment well, these switches allow you to alter certain Trainer parameter settings. See “Configuring the FRx Trainer” for details. (The FRx Defibrillator does not have these switches.
OPTIONAL ACCESSORIES The following accessories are available from Philips separately: • • • • • • • • • • • • 6 PHILIPS MEDIC AL SYSTEMS • FRx Trainer Instructions for Use [REF: 989803139401] HeartStart Training Pads II (kit containing one set of Training Pads II in training pads case, adult pads placement guide, Instructions for Use, and illustrated guide) [REF: 989803139271] Replacement Training Pads II (pair of training pads on disposable liner for use in training pads case provided with HeartStar
SETUP INSTALLING OR REPLACING THE BATTERIES The HeartStart Trainer is powered by four AA alkaline batteries (not included). To install or replace the batteries, follow the steps below: 1. Press the latch on the battery compartment door on the back of the FRx Trainer. Lift off the door and set it aside. 2. Remove the old batteries and recycle or dispose of in accordance with your local regulations. 3. Insert four new AA alkaline batteries into the recess, oriented as illustrated in the battery holder.
REPACKING THE TRAINING PADS II 1. Firmly press one pad into each side of the open training pads case. 2. Untwist the pads wires if necessary. 3. Coil the pads wires inside the case with the connector plug exiting through the hole in the pads case. USING THE MANIKIN ADAPTERS To simulate the patient, use the Training Pads II on a CPR training manikin equipped with a disposable External Manikin Adapter M5089A or an Internal Manikin Adapter M5088A.
INSTALLING THE INTERNAL MANIKIN ADAPTER The Internal Manikin Adapter M5088A, sold separately, works with Laerdal Little Anne CPR and Laerdal Resusci Anne CPR Manikins. To install the Internal Manikin Adapter on a Laerdal Little Anne CPR or Laerdal Resusci Anne CPR Manikin, follow the steps below: 1. Remove the skin from the manikin. 2. Peel off the liner from the adhesive backing of the Internal Manikin Adapter. PHILIPS MEDIC AL SYSTEMS 3.
SELECTING A TRAINING SCENARIO 1. Make sure the pads connector is plugged in and the FRx Trainer is turned off. (Press and hold the On/Off button if necessary to turn off the Trainer.) 2. Press the flashing i-button once. The FRx Trainer will tell you how to select a scenario. 3. Press the i-button again to have the FRx Trainer identify the current training scenario. If you want to use this scenario, briefly press the On/Off button to run the scenario immediately.
4. If using a clothed manikin, remove the clothing as instructed. 5. Open the Training Pads II case and peel off one pad. NOTE: Avoid touching the pad’s adhesive gel, because this may cause the Trainer to begin analysis prematurely. 6. Place the pad on the manikin. Press the pad down firmly. Then peel and place the other pad. PHILIPS MEDIC AL SYSTEMS 7. Continue to follow the FRx Trainer’s instructions for remainder of scenario. 8.
The legend below identifies the symbols used on the rear label of the FRx Trainer and in the following scenario descriptions. NOTE: The shock series can be configured to either one or three shocks. If configured for the default one-shock series, the Trainer provides a pause after each shock. SYMBOL MEANING (Simulated) shockable rhythm detected by Trainer. (Simulated) non-shockable rhythm detected by Trainer. (Simulated) pads problem detected by Trainer.
SCENARIO DESCRIPTION SCENARIO DETAILS Scenario 3 • Trainer detects poor pad contact, repeats pads placement instructions. • After one pad is removed and reapplied, Trainer detects a shockable rhythm, instructs user to deliver a shock. • Trainer detects a non-shockable rhythm.
SCENARIO DESCRIPTION SCENARIO DETAILS Scenario 7 • Trainer detects a shockable rhythm, instructs user to deliver a shock. • Trainer still detects a shockable rhythm, instructs user to deliver another shock. • Trainer detects a non-shockable rhythm. • Trainer detects refibrillation (return to a shockable rhythm), instructs user to deliver a shock. • Trainer detects a non-shockable rhythm.
• • check pad placement, and evaluate user knowledge on topics such as CPR and assessing the patient. TRAINER PERFORMANCE DURING USE PHILIPS MEDIC AL SYSTEMS The Philips HeartStart FRx Trainer is designed for use in an automated external defibrillator training class taught by a qualified instructor. When a training scenario is running, the FRx Trainer mimics the behavior of the FRx Defibrillator during actual emergency use.
CHANGING THE DEFAULT TRAINER CONFIGURATION The HeartStart FRx Trainer comes with a factory default configuration designed to meet the needs of most users. The settings for certain parameters, listed in the table below, can be altered by the user. To change configurable Trainer parameters, remove the battery compartment door from the Trainer and locate the switches in the bottom of the battery compartment. Using a small implement, adjust the ON and OFF switch settings as described in the following table.
PHILIPS MEDIC AL SYSTEMS PARAMETER SWITCH # SETTING (DEFAULT = OFF) CPR Coaching ventilations 5 OFF: ON: with ventilations without ventilations shock series 6 OFF: ON: 1 3 protocol pause* (minutes) 7&8 switch #7 OFF ON OFF ON switch #8 OFF OFF ON ON setting 2.0 1.0 1.5 3.
BEHAVIOR Trainer does not say ANALYZING when pads are applied to the manikin adapter. • Pads placement is incorrect. • Pads are old, damaged, or use discontinued blue gel. • Manikin adapter is broken. • In scenarios 3 and 8, pad has not been removed and repositioned. • Pad wires are twisted. RECOMMENDED ACTION • Check for proper pad and manikin adapter placement. Remove and reposition pads or adapter if indicated. • Replace training pads. • Replace manikin adapter. • Remove and reapply one pad.
ENVIRONMENTAL CONSIDERATIONS By complying with your national regulations regarding disposal of electric, electronic, and battery waste, you can make a positive contribution to our shared environment. Such waste can introduce harmful elements into the environment as a whole and may also endanger human health. PHILIPS MEDIC AL SYSTEMS The FRx Trainer contains electronic components. Do not dispose of it as unsorted municipal waste.
NOTES PHILIPS MEDIC AL SYSTEMS
P H I L I P S M E D I C AL S Y S TE M S INTENTIONALLY BLANK
INTENTIONALLY BLANK P H I L I P S M E D I C AL S Y S TE M S
P H I L I P S M E D I C AL S Y S TE M S INTENTIONALLY BLANK
Philips Medical Systems is part of Royal Philips Electronics Philips Medical Systems United States Philips Medical Systems 2310 Fifth Avenue, Suite 200 Seattle, WA 98121-1825, USA (800) 263-3342 Canada Philips Healthcare, a Division of Philips Electronics Ltd.