Tr ain e r does
not say
ANALYZING
when pads are
applied to the
manikin
adapter.
Pads placement is
incorrect.
•Pads are old,
damaged, or use
discontinued blue
gel.
Manikin adapter is
broken.
In scenarios 3 and 8,
pad has not been
removed and
repositioned.
•Pad wires are
twisted.
Check for proper pad
and manikin adapter
placement. Remove
and reposition pads or
adapter if indicated.
Replace training pads.
Replace manikin
adapter.
Remove and reapply
one pad.
Untwist the pad wires.
Tr ain e r s voi c e
instructions
are distorted.
One of the batteries
is installed
backwards.
Reinstall batteries
according to the
diagram in the battery
holder.
Battery
depletes
prematurely.
Trainer is stored so
that On/Off button
is inadvertently
pressed.
One of the batteries
is installed
backwards.
Store the Trainer so as
to prevent inadvertent
pressure on the On/
Off button.
Reinstall batteries
according to the
diagram in the battery
holder.
BEHAVIOR POSSIBLE CAUSE
RECOMMENDED
ACTION