Philips C3 Patient Monitor C3 Service Guide Models: 862474, 862478 Part Number 989803129451 Printed in the U.S.A.
About this Manual Proprietary Information This document contains proprietary information, which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1085 (978) 687-1501 Publication number 989803129451 Printed in USA Warranty The information contained in this document is subject to change without notice.
Safety All warnings, precautions and notes are located in the chapters that follow. You must read all of this safety information before you begin monitoring with your C3 patient monitor. Safety Standards The C3 Patient Monitor is compliant with the following safety standards: • • • • • • • UL 2601-1 CAN/CSA C22.2 No. 601.
Authorized EU-representative: Philips Medizinsystems Böblingen GmbH, Hewlett Packard Str., 71034, Böblingen Germany Note United States Canada United States Federal Law restricts this device to sale by or on the order of a physician. This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada. Philips Software License Terms ATTENTION: USE OF THE SOFTWARE IS SUBJECT TO THE PHILIPS SOFTWARE LICENSE TERMS SET FORTH BELOW.
Transfer. Your license will automatically terminate upon any transfer of the Software. Upon transfer, you must deliver the Software, including any copies and related documentation, to the transferee. The transferee must accept these License Terms as a condition to the transfer. Termination. Philips Medical Systems may terminate your license upon notice for failure to comply with any of these License Terms.
Text Conventions The following conventions for Notes, Cautions, and Warnings are used in this manual. Warning A Warning calls attention to a condition or possible situation that could cause injury to the user and/or patient. Caution A Caution calls attention to a condition or possible situation that could damage or destroy the product or the user’s work. Note A Note calls attention to an important point in the text. Explanation of Symbols Symbols on products and packaging mean the following.
Keep out of Sun Keep Dry Fragile Keep Upright AC LED Battery LED NBP Volume Contrast On/Standby Snapshot Continuous Defibrillator-proof type CF equipment vii
ECG T SpO2 Temperature SpO2 NBP CO2 Carbon Dioxide Alarm Alarm Limits Menu Audio Off Silence/Reset NBP Automatic Interval Mode Big Numbers Clock Up Alarm Arrow Limit Down Alarm Arrow Limit viii
Empty Battery Perfusion Indicator Bar Heart Rate Powering Down Heart Rate taken from ECG Heart Rate taken from NBP Heart Rate taken from SpO2 Respiration Stat Mode Audio Pause Setup Menu ECG Size Bar RF Interference ix
Battery Compartment RS-232 I/O Equipotential Grounding Post Defib Sync Attention, consult accompanying documents Alternating Current Fuse Type T1AL 250V Date of Manufacture (Y= year; X = Month) YYYY-XX IPX1 Drip-proof ETL Mark x
Sales and Support Offices Please call your local Philips Medical Systems sales office listed in your telephone directory, or a Philips Medical Systems regional office listed below for the location of your nearest sales office or for information on how to contact the Philips Response Center. CORPORATE HEADQUARTERS: Philips Medical Systems Netherlands B.V. Postbus 10.
LATIN AMERICA HEADQUARTERS: Philips Medical Systems 1550 Sawgrass Corporate Parkway #300 Sunrise, FL 33323 Telephone: 954-835 2600 Fax: 954-835-2626 ASIA PACIFIC HEADQUARTERS: Philips Medical Systems 30/F Hopewell Centre 17 Kennedy Road Wanchai Hong Kong Tel: (852) 2821 5888 Fax: (852) 2527 6727 xii
Contents 1. Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3 C3 Patient Monitor Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3. Maintaining the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 Maintenance Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 Objectives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 Concepts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Assurance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11 Power-On Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11 Alarm Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12 Volume Control Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Patient Leakage Current, with Mains Voltage on the Applied Part . . . . . . . . . . . . . . . 4-44 5. Configuring the Power-up Defaults Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 Power-Up Defaults Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Menu Options. . . . . . . . . . .
Removing the Optional, External Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6 Separating the Front from the Rear Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7 Front Case Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8 Replacing the Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11. Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2 Levels of Involvement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2 Preventative Maintenance Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1 Hardware Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-2 Safety Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-2 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents-8
1 Overview This chapter provides a brief overview of the C3 patient monitor. It also provides a list of the monitor’s features and parameters.
General Safety Information General Safety Information Warning This monitor is not intended for neonates. If you connect the monitor to any instrument, verify proper operation before clinical use. Refer to that instrument’s Instructions for Use guide for full instructions. Accessory equipment connected to the monitor’s data interface must be certified according to IEC Standard 60950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment.
Introduction Introduction This manual contains information for servicing the C3 patient monitor, subsequently referred to as "the monitor" throughout this manual. Only qualified service personnel should service this product. Before servicing the monitor, carefully read the C3 Instructions for Use guide for a thorough understanding of operation.
Features and Options Features and Options The measurement parameters and features for each model are indicated below. Order Number Description Display Color Front Panel Description Measurements ECG NBP Temp 862474 Standard Color • • • • • 862478 Sedation Color • • • • • CO2 Recorder (optional) • (optional) The following diagram illustrates the controls located on the front panel of your C3. A B C D E F G Callout 1-4 Overview SpO2 Printout Button/LED A. AC LED B.
Features and Options Front of Monitor Use the following diagram to familiarize yourself with the main features of your C3 monitor. The recorder module and all patient input connectors are located on the monitor’s side panels. C B /min 85 mmHg 1.0mV cm 183/ 107 (149 ) SpO2 180 SpO2 % D 93 A etCO2 CO2 mmHg awRR /min °C T° x X Callout Adult imCO2 33 23 37.8 1 E 01/06/03 01:09:17 Description A. Patient Monitoring Input Connectors B. Main Monitoring Screen C. Handle D.
Features and Options Rear of Monitor The following diagram is of the rear panel. For an explanation of the symbols located on this panel, see “Explanation of Symbols” on page vi. The C3 has four connections on its rear panel. The diagram below is of the rear panel. It shows you how to make the four possible connections (Defib Sync, Equipotential, AC Input, and RS-232 I/O). C A 100-230V~ 50-60 Hz. 1.0A T1AL 250V YYYY-XX 100-230V~ 50-60 Hz. 1.
Features and Options Screen Display The LCD displays parameter values, real-time waveforms, alarm messages and screen selection icons. • Numeric frames displays icons and numeric values of real-time physiologic parameter. • Waveform frames displays real-time waveforms, graphical trend, or tabular trend data. • Message and Icon frames displays alarm messages and screen selection icons. Standard Model /min 85 93 SpO2 % mV 1.
Additional Documentation Sedation Model /min mV 1.0 cm 183/107 (149 ) 180 SpO2 SpO2 % Frames etCO2 CO2 mmHg awRR /min Message and Icon Frames 85 mmHg T °C x X Adult 93 Numeric Frames 33 23 37.8 imCO2 1 01/06/03 01:09:17 Additional Documentation To perform test and troubleshooting procedures and to understand the principles of operation and circuit analysis sections of this manual, you must know how to operate the monitor.
2 Site Preparations This chapter describes how to perform site preparation and how to comply with safety guidelines and requirements.
General Site Preparation Safety Information General Site Preparation Safety Information Warning To avoid contaminating or infecting personnel, the service environment or other equipment, make sure that equipment which has been used before has been appropriately disinfected and decontaminated. Disconnect the monitor from the AC source by unplugging the power cable from the AC power connector located on the rear of the monitor.
Site Preparation Guidelines Site Preparation Guidelines • Check that the environment is reasonably free from dust, vibration, corrosive or explosive gases and is within specified temperature and humidity ranges. See Chapter 12, “Specifications” • Position the monitor so that the display is clearly visible and the controls are easy to reach. Warning, Cautions and Safety Precautions Patient Safety To better secure patient safety, become familiar with the details in Chapter 12, “Specifications.
Preparing to Use the Monitor Preparing to Use the Monitor Warning To avoid contaminating or infecting personnel, the service environment or other equipment, make sure that equipment which has been used before has been appropriately disinfected and decontaminated. Power Source See the section titled, “Electrical” on page 12- 3 for all electrical specifications.
Unpacking the Monitor Disposing of the Monitor To avoid contaminating or infecting personnel, the environment or other equipment, make sure you disinfect and decontaminate the monitor appropriately before disposing of it in accordance with your country’s law for equipment containing electrical and electronic parts. For disposal of parts and accessories, such as ECG and SpO2, where not otherwise specified, follow local regulations regarding disposal of hospital waste.
Returning System Components Returning System Components If it is necessary to return the monitor for service, call the Philips Response Center or your local Philips representative for shipping instructions. Have all equipment serial numbers available when calling. To pack the monitor for return, disconnect all cables. It is not necessary to return sensors, patient cables, NBP tubing and cuff or power cord. Pack the monitor to be returned in its original shipping carton if available.
Mounting Mounting Warning Make sure that you have read all applicable instructions before attempting to install the wall mount. Wall mounts that are intended to support monitors must be capable of supporting 4 times the weight of the monitor when properly installed. Do not mount any portion of a monitoring instrument over a patient’s bed. Do not exceed the maximum rated load specified for each wall mount.
Warnings, Cautions, and Safety Precautions Relating to Wall Mount Installation Warnings, Cautions, and Safety Precautions Relating to Wall Mount Installation 2-8 Site Preparations • Make sure that you have read all applicable instructions before attempting to install the wall mount. • Wall mounts that are intended to support monitors must be capable of supporting four times the weight of the monitor when properly installed.
Mounting the GCX Wall Channel Mounting the GCX Wall Channel To install the GCX Wall Channel and Tilt/Swivel mount, follow the documentation packaged with the channel. Attaching the Mounting Plate to the Monitor Step Action 1 Unpack the mounting plate and check all parts are present. 2 Unscrew the 5 screws from the mounting standoffs located on the rear of the monitor. 3 Place the Wall Plate over the mounting standoffs. 4 Re-insert and tighten the screws to secure the wallplate to the monitor.
Attaching the Monitor to the Wall Channel Attaching the Monitor to the Wall Channel Latch r ito r on s fo t M ip en hil s ati r P tem 3 P d fo ys d s C ture al S Roa ip ic c n il d Ph nufa Me tema 1810 Ma ilips inu A 0 Ph 00 M er, M .A. 30 dov U.S An de in a M 23 01 9 85 00 97 1 IPX Step 2-10 Site Preparations Action 1 Slide the Tilt/Swivel mount into the Wall Plate as illustrated by the above diagram. 2 Make sure the latch snaps into the hole on the upper right-hand corner of the wallplate.
3 Maintaining the Monitor This chapter describes how to clean your monitor and how to keep your monitor in the best working condition.
Maintenance Safety Information Maintenance Safety Information Warning Follow local governing ordinances and recycling plans regarding disposal or recycling batteries and other device components. If the battery shows signs of damage or signs of leakage, replace it immediately. Do not use a faulty battery in the monitor. Never use a monitor that is monitoring a patient to perform battery conditioning. Caution Do not immerse the monitor in liquid or use caustic or abrasive cleaners.
Objectives Objectives In order to meet this chapter’s objectives, you should be able to perform light maintenance and preventative maintenance for the monitor through the following tasks: • Inspection of the monitor • Cleaning the monitor • Battery Maintenance Concepts Light Maintenance Light maintenance tasks can be defined as those tasks required to maintain the monitor in clean, functional working order.
Recommendations for Maintenance Frequency Recommendations for Maintenance Frequency The following table is the recommended maintenance schedule for your C3 monitor. The maintenance checklist appears in the next section of this chapter. The checklist may be photocopied and should be completed by the maintainer. It should be filed for future reference. The owner of the equipment is responsible for the performance of the maintenance activities with the schedule below.
Maintenance Checklist Maintenance Checklist Topics See Page....
Inspecting the Monitor Inspecting the Monitor If you discover a problem while inspecting monitor, contact your biomedical department or Philips Medical Systems. With the monitor turned Off: Step Action 1 Examine the exterior of the monitor for cleanliness and general physical condition. Verify that the housing is not cracked or broken, that everything is present, that there are no spilled liquids and that there are no signs of abuse.
Cleaning Cleaning Caution Do not immerse the monitor in liquid or use caustic or abrasive cleaners. Do not spray or pour any liquid on the monitor or its accessories. Do not allow any liquid to penetrate connectors or openings in the monitor’s chassis. To clean the monitor, dampen a cloth with a commercial, nonabrasive cleaner and wipe the top, bottom, and front surfaces lightly. For cables, sensors and cuffs, follow the cleaning instructions in the Directions for Use shipped with those components.
Battery Maintenance Battery Guidelines To get the most out of the battery, observe the following guidelines: • The shelf-life of a fully charged battery is about 6 months. • After a battery has been activated, keep it charged. If it is not in use, recharge it every 3 to 4 months. • Condition the battery approximately every 200 deep discharges. • Remove the battery from the monitor if it is not being used regularly.
Battery Maintenance Identifying Battery Strength You can determine the monitor’s battery strength by the color of the battery icon, or by the number of illuminated bars on the battery icon.
Battery Maintenance Conditioning a Battery Battery conditioning re-calibrates the battery to ensure that it has accurate information on the actual battery capacity. Warning Never use a monitor that is monitoring a patient to perform battery conditioning. Why is Battery Conditioning Necessary? The capacity of a battery decreases gradually over the lifetime of the battery. Each time a battery is charged its capacity decreases slightly.
Battery Maintenance Battery INOP Messages The following battery-related INOP messages are issued by the monitor. All Battery INOPs continue until the monitor is plugged into the AC power or the INOP condition is fixed. • "Low Battery" - This is an INOP which indicates that the remaining battery-operating time is less than approximately 30 minutes.
Battery Maintenance 3-12 Maintaining the Monitor
4 Testing the Monitor This chapter provides a checklist of testing procedures used to verify performance following repairs or during routine maintenance. All tests can be performed without removing the monitor’s covers. If the monitor fails to perform as specified in any test, repairs must be done to correct the problem before the monitor is returned to the user.
Testing Safety Information Testing Safety Information Warning Perform all leakage tests any time the unit is opened. AC mains voltage is present on the applied part terminals during this test. Exercise caution to avoid electrical shock hazard. Note The monitor must be placed in Diagnostic Mode, with the NBP test screen active for each of the NBP tests. If the test fails, ensure the integrity of the cuff and tubing, then test again.
Objectives Objectives In order to meet this chapter’s objectives, you should be able to test the monitor through the following types of tests: • Performance Assurance Check List • Accuracy, Calibration and Performance Procedures • Patient Safety Checks For inspection procedures; preventative maintenance procedures; cleaning procedures; and battery maintenance, refer to this chapter and Chapter 2, “Site Preparations”.
Testing Checklist Testing Checklist The tests described in this chapter are listed in the table below. Use this table as a checklist. Topics See...
Test Reporting Test Reporting The following table shows what must be recorded on the Service Record after completing the tests in this chapter.
Recommendations for Testing Frequency Recommendations for Testing Frequency The testing checklist appears in the next section of this chapter. Perform the procedure as indicated in the suggested testing timetable. These timetable recommendations do not supersede local requirements. Suggested Testing Frequency Functionality Assurance • Performance Assurance Test When functional defects in the measurements are suspected, after any repairs, if the monitor has been dropped or opened.
Test Map Test Map The Test Map shows which tests are required in which situations. Service Event (When performing...) Test Block Required (...
Equipment Equipment The following table lists the equipment required for performance verification. Equipment 4-8 Testing the Monitor Description Digital multimeter (DMM) Fluke model 87 or equivalent Defib Sync Cable and Switchcraft 3.
Serial Numbers Serial Numbers When recording test results, these are always associated with a particular monitor by means of the serial number. The serial number is 10 characters and is located on the back of the monitor. Serial Number 100-230V~ 50-60 Hz. 1.0A T1AL 250V USC3xxxxxx YYYY-XX CE 0123 PHILIPS C3 PATIENT MONITOR Manufactured for Philips for Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 Made in U.S.A.
Visual Test Visual Test Inspect the system for obvious signs of damage. Also check external leads and accessories. What to record on the service record: V:P or V:F Power On Test Step Action 1 Switch the monitor On. 2 Observe whether the system boots up successfully without displaying any error codes. The main monitoring screen should display. What to record on the service record: PO:P or PO:F .
Functionality Assurance Tests Functionality Assurance Tests The following assurance checks are recommended to verify proper operation daily before the monitor is used to monitor a patient. During functionality assurance checks, verify the overall operation by completing the following Performance Assurance Tests. Performance Assurance Test To verify your monitor works properly, perform the following test: Step Action 1 Using the supplied AC power cord, connect the monitor to the AC power source.
Functionality Assurance Tests Alarm Test Tools Needed: • SpO2 adapter cable (M1943A) • Clinical Dynamics Corp - SmartSat simulator (with Nellcor simulator cable) Step Action 1 Connect SpO2 simulator cable to the SpO2 adapter cable. Connect the cable to the SpO2 patient monitoring input connector. 2 Set the simulator as follows: Item Setting Oximeter Nellcor SpO2% 81 BPM 36 Pulse Mod 0.50% 3 Press the On/Standby button to turn the monitor On.
Functionality Assurance Tests Step Action 1 Connect SpO2 simulator cable to the SpO2 adapter cable. Connect the cable to the SpO2 patient monitoring input connector. 2 Set the simulator as follows: Item Setting Oximeter Nellcor SpO2% 81 BPM 70 Pulse Mod 5.00% 3 a. Power the monitor On. Verify that the values are within the following tolerances: – Oxygen Saturation Range 79% to 83% – Heart Rate Range 67 to 73 bpm 4 Press the Alarm Silence button to temporarily silence the audible alarm.
Preventative Maintenance Tests Preventative Maintenance Tests Preventative maintenance refers specifically to the service tests required to make sure the monitor measurement results accurate. In cases where the performance of NBP is in question or could have been configured during repair, the complete set of NBP tests described in this service manual should be used. The tests in this section verify the functionality of the monitor’s pneumatic system.
Preventative Maintenance Tests Documenting NBP Test Results The following table lists the tests that should be documented and summarizes how to document the NBP test results. Test Expected Test Results Pressure Transducer Accuracy Test X1 = difference between the pressure on the monitor and on the simulator Difference less than or equal to 5 mmHg (0.
Preventative Maintenance Tests Pneumatic System Functionality The following tests must be performed to verify pneumatic system functionality. Perform these tests in the following order: 1. 2. 3. 4. 5. Pressure Transducer Accuracy Pneumatic Leakage Inflation Rate Over-Pressure Deflation Rate Note The pneumatic system includes an over-pressure safety limit function and a safety period time-out function. These safety functions may interfere with NBP tests described in this section.
Preventative Maintenance Tests Pressure Transducer Accuracy Test This test verifies the pressure accuracy of the monitor’s pressure transducer. Step Action 1 Confirm that the simulator displays "Leak Test" (see steps 1 and 2, in the previous table). 2 Press the Volume button to ensure that both valves are closed. 3 Perform an offset adjustment so that the simulator and monitor both display a pressure of 0 mmHg or kPa by doing the following: a.
Preventative Maintenance Tests Step Inflation Rate Test Action 8 Start a one minute timer. After one minute, record the pressure displayed again (P2). The pressure should drop by no more than 6 mmHg (0.8 kPa) during the one minute period. Calculate (P1 - P2) and document the leakage test value (X2). 9 Press and hold the Volume button until the monitor displays a pressure of 0 mmHg or kPa. 10 If no further NBP tests are to be conducted, turn the monitor Off.
Preventative Maintenance Tests Over-Pressure Test This test verifies the functionality of the over-pressure relief system of the monitor. Step Action 1 Turn the simulator on. 2 Perform the following sequence: a. Press the Home button. b. Press the Pressure Tests button. c. Press the Pressure Relief button. d. Press the Setup button. e. Press 3, 3, 5, then Enter. 3 Confirm that the simulator is active and displays "Relief Valve Test".
Preventative Maintenance Tests Deflation Rate Test This test verifies the deflation rate of the monitor. Step Action 1 Turn the simulator on. 2 Perform the following sequence: a. Press the Home button. b. Press the Pressure Tests button. c. Press the Pressure Leak Test button. d. Set the Cuff to Internal. 3 Confirm that the simulator is active and displays "Leak Test". 4 Press the Volume button to ensure that both valves are closed.
Preventative Maintenance Tests Basic Pneumatic Leakage (BPL) Test The purpose of this test is to verify the integrity of the NBP pneumatic system after the monitor has been opened. This includes all external and internal tubing connections. No simulator is required for this test, and results are not required to be reported. Step Action 1 Attach the NBP cuff to the NBP tubing and the tubing to the NBP patient monitoring input connector.
Performance Procedures Performance Procedures The following accuracy and performance procedures are designed to be completed to verify the accuracy and performance of the monitor. They must be performed, according to the frequency specified in the section titled, “Recommendations for Testing Frequency” on page 4-6. Battery Performance Test The battery performance test should be performed every two years, before monitor repairs, or whenever the battery is suspected as being a source of the problems.
Performance Procedures Step Action 5 With the monitor Off, press the On/Standby button. (When the monitor is turned on, the battery icon may initially indicate a higher charge then it actually holds wait 5 minutes or until after an NBP measurement for an accurate battery reading.) 6 Verify that the monitor is responding to the SpO2 simulator signal and that the audible alarm is sounding. Use the navigation wheel to select the SpO2 Menu and permanently silence the SpO2 audible alarm.
Performance Procedures Step 3 Action Set the temperature simulator as follows: a. Temperature: 37°C (98.6°F). b. Probe Type: YSI 400 Series 4-24 Testing the Monitor 4 Press the On/Standby button to turn the monitor On. 5 After the power-up sequence, verify the temperature reads 37°C ±0.1°C (98.6°F ±0.2°F if Fahrenheit is selected as the temperature unit). 6 Turn the monitor Off.
Performance Procedures ECG/ Respiration Performance ECG Test This section includes tests for both ECG and Respiration. Tools Needed for both ECG and Respiration testing: • ECG leads - M1605A (AAMI) or M1615A (IEC) • ECG trunk cables - M1540C (ICU, AAMI) or M1550C (ICU, IEC) • Dynatech Nevada MedSim 300 or equivalent simulator The accuracy of the monitor’s ECG measurements is ±5 bpm. In the following procedure, add the tolerance of the simulator to the acceptable range of readings.
Performance Procedures Respiration Test The accuracy of the monitor’s respiration measurements is ±3 breaths per minute. In the procedure below, add the tolerance of the simulator to the acceptable range of readings. Step Action 1 Verify that the monitor is turned Off. Connect the ECG leads to the appropriate jacks on the ECG simulator. 2 Connect the ECG leads to the ECG trunk cable. 3 Connect the cable to the ECG patient monitoring input connector.
Performance Procedures SpO2 Performance SpO2 testing includes the following tests: • Dynamic Operating Range • LED Excitation Test Tools Needed for SpO2 Tests: Dynamic Operating Range Test • Clinical Dynamics Corp - SmartSat simulator (with Nellcor simulator cable) • SpO2 adapter cable (M1943A) • SpO2 reusable sensor, adult finger (M1191A) The following test sequence verifies proper monitor operation over a range of input signals.
Performance Procedures LED Excitation Test This procedure uses normal system components to test circuit operation. A SpO2, adult finger, reusable sensor is used to examine LED intensity control. The red LED is used to verify intensity modulation caused by the LED intensity control circuit. Step 4-28 Testing the Monitor Action 1 Connect an SpO2 sensor to the monitor. 2 Press the On/Standby button to turn the monitor On.
Performance Procedures CO2 Performance This test checks the performance of your CO2 measurement for the monitor. This test uses calibration equipment that can be ordered (contact your Philips representative). Refer to the documentation accompanying the equipment for detailed instructions.
Performance Procedures Documenting CO2 Test Results The following table lists the CO2 tests and summarizes how to document the test results.
Performance Procedures Barometric Pressure Check and Calibration Check the barometric pressure value in the CO2 module as follows: Step Action 1 Enter the Power-Up Defaults Menu (see ““Power-Up Defaults Menu” on page 5-3“for instructions on accessing this menu). 2 Select the Enter Diagnostic Mode? option. 3 From the popup menu, select Yes. The Diagnostic Menu displays. 4 Select the CO2 Test option. 5 Connect a FilterLine to the CO2 patient monitoring input connector.
Performance Procedures Leakage Check The Leakage Check consists of two parts: • Part 1 - Checking the tubing between the pump outlet and the CO2 module outlet. • Part 2 - Checking the tubing between the pump inlet and the FilterLine inlet. Note Check the flowmeter’s user guide for details on how to make a correct flow reading. Part 1 Step Action 1 Check the ambient and cell pressure shown in the status line on the screen. The cell pressure should be approximately 20 mmHg (2.
Performance Procedures Pump Check Step Action 1 Connect the flowmeter inlet to the CO2 module gas outlet. 2 Connect the FilterLine to the CO2 module inlet. 3 Block the inlet of the FilterLine using your fingertip and observe the cell pressure to the right of the slash symbol on the top row of the CO2 Test display screen. The cell pressure (X4) should be more than 120 mmHg below the ambient pressure shown.
Performance Procedures CO2 Gas Measurement Calibration Check After switching the monitor on, or after turning on the etCO2 On/Off setting in the etCO2 frame, wait at least 20 minutes before checking the calibration. Step Action 1 Check that the 5% calibration gas and flow regulator are connected. 2 Calculate the expected measurement value in mmHg as follows: 0.05 x (ambient pressure) / 1.03 = value mmHg i.e. 0.05 x 736 mmHg / 1.03 = 35.
Performance Procedures Calibration Verification Step Reset Pump Operating Time Counters Action 1 Reopen the 5% gas valve and allow the value to stabilize. 2 Check that the value displayed on the monitor is correct and within the tolerance (see Step 2 in above section). 3 Disconnect the 5% calibration gas and connect the 10% calibration gas. 4 Open the valve on the flow regulator to allow 10% CO2 gas to flow into the monitor. Allow the value to stabilize.
Performance Procedures Serial Interface and Nurse Call Signal Test Perform the following procedure to test the serial port voltages. The test is qualitative and only verifies that the serial interface port is powered correctly, and that the Nurse Call signal is operational. The serial connector is a male DB-9 located on the monitor’s rear panel, identified by the RS-232 symbol.
Performance Procedures Pin Signal Direction Measurement (V) 8 CTS<<< input -0.4 0.0 0.4 9 Alarm Out>>> output -5.0 -7.0 -15.0 output 5.0 7.0 15.
Performance Procedures ECG Sync Test This test checks the performance of ECG synchronization between the monitor and a defibrillator. The ECG sync performance test is required once every year and when the monitor is repaired or when the monitor’s parts are replaced. Tools Needed: • Dynatech Nevada MedSim 300 or equivalent simulator • Defib Sync Cable (M4820A) or Switchcraft 850 • Switchcraft 750 Two sections are needed to complete the connection from the monitor to the simulator. 1.
Patient Safety Tests Patient Safety Tests Philips safety tests meet the standards of, and are performed in accordance with IEC 60601-1, Clause 19 (EN60601-1, Second Edition, 1988; Amendment 1, 1991-11, Amendment 2, 199503). The C3 patient monitor is a Class I device. It requires a protective earth (ground) wire. Keep this in mind when performing the following test procedures. There are two categories of safety tests: Ground Integrity • Ground Integrity • Electrical Leakage Table 4-3.
Patient Safety Tests Figure 4-1 Safety Test Diagram - Protective Earth Test or Inspection to Perform Measures impedance of Protective Earth (PE) terminal to all exposed metal parts of IUT, which are for safety reasons connected to the Protective Earth (PE). Max. 100 mOhm. Test current 25 Amps applied for 5 to 10 seconds. We recommend to flex the main cable during the test in order to identify potential bad contact or damage of the earth wire. Safety test according to EN60601-1 (Clause 18).
Patient Safety Tests Electrical Leakage The following tests verify the electrical leakage of the monitor: • Earth Leakage Current • Enclosure Leakage Current • Patient Leakage Current • Patient Source Current, with Mains Voltage on the Applied Part Warning Perform all leakage tests any time the unit is opened. Earth Leakage Current This test is in compliance with IEC 60601-1 (Earth Leakage Current). In locations where mains voltage is 100-120 volts, the applied voltage is 132 volts.
Patient Safety Tests Enclosure Leakage Current This test is compliance with EN60601-1 (Enclosure Leakage Current). Test at 110% of the nominal line voltage. Step Action 1 Connect the AC mains power cord to the analyzer as recommended by the analyzer operating instructions. 2 Using the appropriate test cable, connect the analyzer to either of the screws on the back of the monitor, next to the handle. 3 Turn the monitor on. 4 Perform the test as recommended by the analyzer operating instructions.
Patient Safety Tests Patient Leakage Current This test measures patient leakage current in accordance with EN60601-1, Clause 19, for Class I, type CF equipment. Patient leakage current in this test is measured from any individual patient connection to earth (power ground). This test requires a sample patient cable for each device parameter. These must be configured as recommended by the safety analyzer operating instructions.
Patient Safety Tests Patient Leakage Current, with Mains Voltage on the Applied Part This test measures patient leakage current in accordance with EN60601-1, Clause 19, for Class I, type CF equipment. In this test, 110% of mains voltage is applied between each patient connection and earth (power ground). Patient leakage current is then measured from any individual patient connection to earth. Warning AC mains voltage is present on the applied part terminals during this test.
Patient Safety Tests Testing the Monitor 4-45
Patient Safety Tests 4-46 Testing the Monitor
5 Configuring the Power-up Defaults Menu This chapter discusses the use of the Power-up Defaults Menu to configure power-on default settings. It also explains how to use Diagnostic Mode to obtain service-related information about the monitor.
General Safety Information General Safety Information Warning A blood pressure cuff, connected to the monitor, should never be applied to a human being while the monitor is in Diagnostic Mode. Injury could result. A FilterLine, connected to the monitor, should never be applied to a human being while the monitor is in Diagnostic Mode. Injury could result. Caution In addition to restoring factory defaults, this procedure clears the contents of trend memory.
Power-Up Defaults Menu Power-Up Defaults Menu The purpose of the Power-up Defaults Menu is to allow the authorized user to create a power-up default for each setting in the monitor. Power-up defaults are the settings in effect each time the monitor is power on. After you enter the Power-up Defaults Menu, physiological monitoring is terminated. The screen layouts do not display any information associated with normal monitoring operation.
Power-Up Defaults Menu Menu Options Below are the menu options you can configure from the Power-up Defaults Menu. Menu Item Choices Explanation Accept Current Settings Yes If Yes is chosen, the current monitor settings become the power-up defaults. No Selected Audio Off Make Available (default) Deny Access If Make Available is chosen, audible alarms are permanently silenced for a particular parameter via the Alarm Limits Menu.
Diagnostic Mode Diagnostic Mode The purpose of the Diagnostic Mode Menu option is to allow factory, field-service, and hospital biomedical technicians access to a series of test and system-related information screens for the purpose of verifying monitor performance, or troubleshooting problems. To access Diagnostic Mode: Step Action 1 Enter the Power-Up Defaults Menu (see ““Power-Up Defaults Menu” on page 5-3“for instructions on accessing this menu). 2 Select the Enter Diagnostic Mode? option.
Diagnostic Mode Error Codes DIAGNOSTIC MENU ERROR CODES Error Codes 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. Date/Time Return This screen displays the 10 most recent error code types, logged by the monitor. The date and time that the error code appears displays to the right of the error code. Error codes cannot be changed or reset in this screen. When in the Error Code screen, the Return option is always highlighted. Press the wheel to return to the main monitoring screen.
Diagnostic Mode Menu Item Function Monitor On-Time Displays the number of hours, rounded to the nearest hour, that the Main PCB has been operational. This value CANNOT be reset. Backlight On-Time Displays the number of hours, rounded to the nearest hour, that the LCD Backlight has been operational. This value can be reset to zero. An example is when a technician changes the backlight or installs a new LCD.
Diagnostic Mode System A/D Values The System A/D screen displays the current value of each analog-to-digital (A/D) channel in volts. Some of the channels are for AC-coupled signals (such as ECG input), so the numbers on the screen are constantly changing when an input signal is present. These AC-coupled values are shown to give an indication as to whether basic functionality of the channel is present, but no significance can be derived from the values of the numbers displayed.
Diagnostic Mode Table 5-1. A/D Channel Designators 1. ECG 2. RWAVE 3. PACEMAKER 4. RESPIRATION 5. PRESSURE XDUCER 1 6. PRESSURE XDUCER 2 7. NBP OSCILLATORY 8. ECG LEAD OFF 9. TEMPERATURE 10. ISOLATED VOLTAGE REF 11. ISOLATED VOLTAGE ZERO 12. SpO2 S1 S018 13. NOT USED 14. +3.3 VDC POWER SUPPLY 15. (+12 VDC POWER SUPPLY) x 0.33 16. (NBP VOLTAGE REF) x 0.8 17. GROUND REFERENCE 18. (+5 VDC POWER SUPPLY) x 0.8 19. ADC MID-SCALE VALUE 20. ADC ZERO-SCALE VALUE 21.
Diagnostic Mode NBP Test A NBP Test screen is provided to facilitate troubleshooting problems and performing verification testing for the NBP subsystem. Typically, when these tests are performed, the pneumatic system is connected to an external pressure-reading device and a closed reference volume. The NBP Test screen provides a real-time numeric display of the pressure in the pneumatic system, means for controlling the pump and valves are open or closed.
Diagnostic Mode Menu Item Function Activate Pump As long as the NBP button is pressed, the pump runs. If system pressure reaches the hardware over-pressure protection point (280 to 330 mmHg or 37.3 to 44.0 kPa), the safety valve opens and the pump disables, until the pressure falls below the safety threshold. Open Both Valves As long as the Volume button is pressed, the safety valve and proportional valve opens and remains at maximum.
Diagnostic Mode CO2 Test The CO2 Test Screen is provided to facilitate troubleshooting problems and performing verification testing to ensure the measuring results are accurate. See “CO2 Performance” on page 4-29 for information on how to perform these tests. Warning A FilterLine, connected to the monitor, should never be applied to a human being while the monitor is in Diagnostic Mode. Injury could result.
Diagnostic Mode The CO2 Test Screen is described below: Menu Item Function CO2 Pressure in mmHg (ambient/cell) Displays the real-time value of the ambient pressure and the measured cell pressure. The ambient pressure value will be the same as the Barometric Pressure value. This value CANNOT be changed or reset. Pump Voltage (Volts) Takes the correct amount of air from outside of the monitor. This value CANNOT be changed or reset. CO2 (mmHg) Displays the real-time value of CO2.
Restoring Factory Settings Restoring Factory Settings Caution In addition to restoring factory defaults, this procedure clears the contents of trend memory. The following technique can be used to restore the monitor’s power-up default settings which were originally configured in the factory. Note Before performing the first step, read this procedure COMPLETELY. Step Action 1 Ensure that the monitor is powered Off. 2 Simultaneously press the Volume and Contrast buttons on the monitor’s front panel.
6 Troubleshooting This chapter explains how to troubleshoot the monitor if problems arise. Tables are supplied that list possible monitor difficulties, along with probable causes, and recommended actions to correct the difficulty. Use this chapter when you suspect you have a problem with your monitor, or whenever you are referred to this chapter from another part of the Service Guide.
General Troubleshooting Safety Information General Troubleshooting Safety Information Warning Do not attempt to troubleshoot the monitor if there is any sign of burning or smoking coming from the monitor. Call the Philips Response Center or your local Philips representative immediately. Caution Only qualified service personnel should open the monitor housing, remove and replace components, or make adjustments.
Objectives Objectives In order to meet this chapter’s objectives, you should be able to diagnose and isolate hardware failures to the level of the "field replaceable part". In addition, this chapter describes how to troubleshoot problems using error codes. Caution Only qualified service personnel should open the monitor housing, remove and replace components, or make adjustments.
Part 1 Troubleshooting Checklist Part 1 Troubleshooting Checklist If any unit is not functioning properly, your first step is to Restore Factory Settings. See “Restoring Factory Settings” on 5-14 for additional information. Warning Do not attempt to troubleshoot the monitor if there is any sign of burning or smoking coming from the monitor. Call the Philips Response Center or your local Philips representative immediately.
Part 1 Troubleshooting Checklist The following two steps are to make sure that the monitor’s AC power supply and battery work correctly. What To Do If The Monitor Cannot Be Switched On, AC Powered: Step Action 1 Ensure the battery has had ample time to charge (at least 15 minutes). A dead battery will prevent the monitor from powering On. 2 Check AC fuses in the AC receptacle. 3 Check keypad connector or replace keypad. See “Removing the Keypad on the Front Panel” on 7-6 for instructions.
Part 2 Isolating and Solving Monitor Problems Part 2 Isolating and Solving Monitor Problems INOP Messages The following table explains the technical INOP messages that the monitor can issue, and suggests a course of actions. When actions are numbered, always try them in the order given and only proceed to the next action if the current one is not successful in solving the problem. Message Cause of Failure Remedy "CO2 Equip Malf" Malfunction in the CO2 hardware.
Part 2 Isolating and Solving Monitor Problems Message Cause of Failure Remedy "Loss of Pulse from SpO2" SpO2 cannot find a pulse. Make sure the SpO2 patient cable is connected and the sensor is correctly attached to the patient. "Low Battery" The battery has less than 30 minutes of charge left. Recharge the battery. "NBP - Artifact" Incomplete measurement due to patient motion or physiology. "NBP - Blocked Hose" NBP tubing blocked - measurement not possible.
Part 2 Isolating and Solving Monitor Problems Message Cause of Failure Remedy "SpO2 Noisy Signal" Excessive patient movement or electrical interference are causing irregular pulse patterns. Try to reduce patient movement, or to relieve the cable strain on the sensor. "SpO2 Non-Pulsatile" Pulse is too weak or is not detectable, or the application site is too thin. Change the application site of the sensor, or stimulate circulation at the current site. "SpO2 Transduc Malf" Sensor is malfunctioning.
Part 2 Isolating and Solving Monitor Problems Symptom Cause of Failure Remedy Monitor fails to power-up when the On/Standby button is pressed. Battery is dead or monitor is not plugged in. Make sure the monitor is powered on by either battery or AC power. Check that the battery is adequately charged and AC power fuses are securely connected. Keypad malfunction or keypad connector is loose. 1. Check the keypad connector. 2.
Part 2 Isolating and Solving Monitor Problems Symptom Cause of Failure No sound from monitor Tone is switched off or low. Incorrect Time Stamps The monitor’s real time clock is defective. Power the monitor On and then reset the time and date. See the C3 Instructions for Use guide for instructions on how to set the date and time. Cannot reprogram the time and date correctly Main PCB module malfunction.
Part 3 Using Support Functions Part 3 Using Support Functions When the monitor detects an error condition, the monitor shows an error code on the display screen. If such an error occurs during monitoring operation, an audible alarm tone also sounds. Press the Alarm Silence button to terminate the audible alarm tone. When an error code appears on the display, a number in hexadecimal representation indicates the nature of the error.
Part 3 Using Support Functions Hex Code Error Code Categories 6-12 Troubleshooting Explanation Recommended Action 64 The SpO2 module is sending an error message to the host CPU. 1. Ensure PicoSat module is properly connected. 2. Replace the PicoSat module. 3. If problem persists, replace main PCB module. ? The CO2 module is sending an error message to the host CPU. 1. Ensure Oridion module is properly connected. 2. Rplace the Oridion module. 3. If problem persists, replace main PCB module.
Part 3 Using Support Functions Other Error Codes Hexadecimal Code Explanation 514xxx system error 515xxx CO2 error If an error code occurs that is not listed in the above sections, take the following actions: Step Action 1 Power cycle the monitor. 2 If the error code still displays, take the monitor out of service and contact the Philips’ Response Center or your local Philips representative for advice on remedial action.
Part 3 Using Support Functions 6-14 Troubleshooting
7 Disassembly This chapter provides step-by-step procedures that are used to access replaceable parts of the monitor. The sections in this chapter describe and photographically illustrate procedures for disassembling the monitor in order to remove or replace suspected defective assemblies or components.
Introduction Introduction The monitor consists of two main assemblies, the front case assembly, and the rear case assembly. All part numbers and exploded views of some assemblies are located in Chapter 8, “Spare Parts” on page 8-1.
Disassembly Safety Information Disassembly Safety Information Warning Performance Verification: Do not place the monitor into operation after repair or maintenance has been performed, until all recommended Performance and Safety Tests listed in Chapter 4, “Testing the Monitor” of this Service Guide have been performed. Failure to perform all tests could result in erroneous monitor readings. Before attempting to open or disassemble the monitor, disconnect the power supply from the monitor.
Disassembly Procedures Disassembly Procedures This section provides instructions for closed case, front case and rear case disassembly procedures. Closed Case Disassembly Procedures This section describes the items that can be removed without disassembling the main case of the monitor. Removing the Battery Warning Before you begin any disassembly procedure, you must remove the battery.
Disassembly Procedures Step Action 3 Disconnect both battery lug terminals to release the battery. 4 When you re-insert the battery, verify that the battery is inserted the correct way and the polarity is correct (see figure below). Black Red Removing the Navigation Wheel Step Action 1 Firmly grasp both sides of the wheel and pull straight back from the monitor. The wheel should slip off the encoder shaft.
Disassembly Procedures Removing the Keypad on the Front Panel Removing the Optional, External Recorder 7-6 Disassembly Step Action 1 The keypad is attached with an adhesive to the front panel. To remove it, carefully lift up one corner of the keypad with a flathead screwdriver. 2 Peel the adhesive away from the front panel. Step Action 1 Press the external Paper Eject button on the right hand side of the recorder. The door drops forward. 2 Remove the paper roll.
Disassembly Procedures Separating the Front from the Rear Case Step Action 1 Verify that the battery has been removed. If not see, “Removing the Battery on page 7-4. 2 Verify that the navigation wheel and brass hex nut have been removed. If not see, “Removing the Navigation Wheel on page 7-5. 3 Use a Philips no. 2 screwdriver to remove the 6 screws fastening the Rear Case Assembly to the Front Case Assembly.
Repeater Disassembly Procedures Step 5 Action Disconnect the display cable located on your left-hand side by pulling the connector towards you. Disconnect the backlight tube, front panel and speaker cables located on your right-hand side. Front Panel Cables Backlight Tube Cables Speaker Cable 6 Front Case Disassembly Display Cable Separate the front from the rear case.
Disassembly Procedures Replacing the Speaker Step Replacing the Backlight Tube Action 1 Use a Philips no. 2 screwdriver to remove the 3 screws fastening the speaker. 2 Pull the speaker up. The backlight tube is located within the LCD screen. However, you do not need to disassemble the LCD screen display to replace the backlight tube. Step Action 1 Insert a small, thin, flathead screwdriver into the crevice between the LCD screen and where the backlight tube slides in.
Disassembly Procedures Rear Case Disassembly You must disassemble the rear case to remove the following modules: • Main PCB • Patient Monitoring I/O • SpO2 • CO2 • NBP • Communications • Power Supply Caution Before you remove any module, make sure you are wearing ESD protection and you are working in a grounded environment. To disassembly the rear case: Step 7-10 Disassembly Action 1 Verify that the battery has been removed. If not see, “Removing the Battery on page 7-4.
Disassembly Procedures Step 3 Action Use a Philips no. 2 screwdriver to remove the 9 screws fastening the rear case. Screws 1-5 Screws 6-9 4 Remove the two screws located in the bottom corners of the rear case. Screw 1 5 Screw 2 Gently slide the chassis away from the rear case.
Disassembly Procedures Removing the Main PCB Module Step 1 Action Unscrew the PCB module by loosening the following 3 screws: Screw #1 Screw #2 Screw #3 2 Gently pull the module towards you. 3 After you have removed the Main PCB module, you should have access to the CO2, SpO2, NBP pump, and the Patient Monitoring I/O modules.
Disassembly Procedures Removing the Patient Monitoring I/O Module Follow these instructions for removing the I/O module from the Main PCB module. Step 1 Action Remove the following 2 screws to remove the connector module. Screw #1 Screw #2 Note—Use a flathead screwdriver to remove Screw #1. 2 Slide the I/O module out. 3 Disconnect the CO2 connector and tubing and the NBP tubing. Slide the CO2 exhaust tubing out. 4 Pull the I/O module away.
Disassembly Procedures Removing the SpO2 Module After you have removed the Patient Monitoring I/O module from the Main PCB module, you can remove the SpO2 module. To do this: Step 1 Action Remove the 2 screws fastening the SpO2 module to the I/O module. Screw #2 Screw #1 7-14 Disassembly 2 Remove the connector that attaches the SpO2 module to the I/O module. 3 Gently pull the SpO2 module away.
Disassembly Procedures Removing the CO2 Module Note—This procedure is only for those that have the C3 Sedation model. Step 1 Action Remove the 4 screws on the CO2 module. Screw #1 Screw #2 Screw #3 Screw #4 2 Tilt the CO2 module towards the NBP pump. 3 Gently pull the module away from the Main PCB module. Note—When reassembling the CO2 module, ensure that the CO2 header matches up with the mating header.
Disassembly Procedures Removing the NBP Pump Step 1 Action Remove the connector from the NBP pump. Connector 2 Cut the tie wrap.
Disassembly Procedures Removing the Power Supply Module Step 1 Action Use the needle-nose pliers to unplug the three connectors. Battery Cables Connector #1 Power Supply Board Connectors #2 & 3 Communications Module 2 Remove the two screws that connect the power supply to the rear panel. 3 Remove the screws on the side and bottom of the rear panel. 4 Cut the tie wrap that holds the battery cables together. 5 Use a 5/16 nut driver to unfasten the power supply from the rear panel.
Disassembly Procedures 7-18 Disassembly
8 Spare Parts This chapter lists spare parts, along with part numbers, are listed in the tables that follow. Item No. corresponds to the callout number in Figure 8-1 through 8-6. the Page Ref. for Disassembly indicates the page number of the relevant disassembly procedures in Chapter 7, “Disassembly”.
Small Parts Kit Small Parts Kit Most small parts and hardware (screws, hex nuts, washers, standoffs, 2 fuses, CO2 spring and door, and 2 housing feet) can be found in the Small Parts Kit, Part Number 453563480201. The picture below calls out a few of the less commonly known parts within this kit.
C3 Top Level Assembly C3 Top Level Assembly Rear Casing Front Casing Figure 8-1 Spare Parts 8-3
Front Panel Assembly Front Panel Assembly 3 1 Figure 8-2 (1 of 2) 8-4 Spare Parts Description Page Ref. for Disassembly Item No.
Front Panel Assembly 8 3 7 4 6 5 Figure 8-3 (2 of 2) Description Page Ref. for Disassembly 453563480021 Backlight Tube page 7-9 453563480101 Keypad page 7-6 5 453563480221 Speaker (including speaker cable) page 7-9 6 453563480081 Display Cables page 7-7 7 453563480051 Bezel Protective Window Display No Disassembly Procedure 8 453563480091 Front Housing No Disassembly Procedure Item No.
Rear Panel Assembly Rear Panel Assembly 9 12 14 Red to (+) Terminal Black to (-) Terminal 10 13 11 Figure 8-4 (1 of 3) 8-6 Spare Parts Item No. Part Number Description Page Ref.
Rear Panel Assembly 15 16 17 Figure 8-5 (2 of 3) Part Number Description Page Ref. for Disassembly 15 453563480191 Side Panel - with CO2 page 7-15 Not Shown 453563480181 Side Panel - without CO2 No Disassembly Procedure 16 453563480121 NBP Assembly page 7-16 16 453563480131 NBP Cable page 7-16 17 453563480141 Patient Monitoring I/O Module page 7-13 Item No.
Rear Panel Assembly 18 Figure 8-6 (3 of 3) Part Number Description Page Ref. for Disassembly 18 453563480161 Power Supply Board page 7-17 Not Shown 453563480061 Communication Board page 7-17 Item No.
Power Cords Power Cords Part Number Description 8120-5429 Power Cord, US (903) 2.4 m 8120-1689 Power Cord, Europe (902) 2.0 m 8120-1351 Power Cord, UK (900) 2.3 m 8120-4475 Power Cord, Australia (901) 2.0 m 8120-2104 Power Cord, Switzerland (906) 2.0 m 8120-2956 Power Cord, Denmark (912) 2.0 m 8120-4211 Power Cord, South Africa/India (917) 2.0 m 8120-5342 Power Cord, Japan (918) 2.5 m 8120-5182 Power Cord, Israel (919) 2.0 m 8120-6869 Power Cord, Argentina (920) 2.
Exchange Unit Part Numbers Exchange Unit Part Numbers 8-10 Spare Parts Exchange Unit Part Number Description 453563344311 Thermal Recorder for C3 Patient Monitor GSI 453563480441 Exchange Recorder for C3 Patient Monitor GSI 453563480381 C3 Standard Color Monitor 453563480401 C3 Standard Color Monitor with Recorder 453563480411 C3 Sedation Color Monitor 453563480431 C3 Sedation Color Monitor with Recorder
9 Packing for Shipment This chapter explains how to pack the monitor for shipment for any reason.
General Instructions General Instructions To ship the monitor for any reason, follow the instructions in this section. To pack the monitor for return, disconnect all cables. It is not necessary to return sensors, patient cables, NBP tubing and cuff or power cord. Pack the monitor carefully. Failure to follow the instructions in this section can result in loss or damage not covered by the warranty. If the original shipping carton is not available, use another suitable carton.
10 RS-232 Interface This chapter explains how to properly use and connect the RS-232 interface and cables.
General RS-232 Safety Information General RS-232 Safety Information Warning If you connect the monitor to any instrument, verify proper operation before clinical use. Refer to the device’s manual for full instructions. Accessory equipment connected to the monitor’s data interface must be certified according to IEC Standard 950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 60601-1 system requirements.
About the RS-232 Interface About the RS-232 Interface The RS-232 interface allows you to : • send a nurse call signal • export trend data to an external PC Warning Anyone who connects additional equipment to the signal input port or signal output port configures a medical system and is therefore responsible to ensure that the system complies with the requirements of system standard IEC Standard 60601-1-1. If in doubt, consult Philips Medical Systems Response Center or your local Philips representative.
About the RS-232 Interface Pin connections for the 9-pin connector are as follows: Table 10-1. RS-232 Serial Interface Connections Nurse-Call Pin # Signal Direction 1 not used 2 Rx data input 3 Tx data output 4 DTR output 5 Signal Ground input/output 6 DSR input 7 RTS output 8 CTS input 9 Alarm Out output Pin 9 of the RS-232 serial interface connector provides an Alarm Out signal.
11 Training This chapter provides a complete, flexible training program that addresses the needs of technical personnel who service the monitor at all levels of involvement.
Introduction Introduction The training itself is self-paced. This chapter guides you through the entire service training program. All of the information you require is in this chapter, or referred to in this chapter.
Training Materials Training Materials Essential Materials The essential materials to complete this training are the following: • The C3 Service Guide (989803129451) For Philips Service Personnel, a Training Completion Form must be signed and returned to your supervisor referencing the course number "MWPMD-C3SUPP".
Support Strategies Support Strategies This section provides a summary of how the monitor was designed to be supported and how Philips provides that support. There are two main methods of repairing the monitor: • Unit exchange • Bench repair Unit Exchange The primary repair method used by Philips service personnel is unit exchange. The current design of the monitor requires testing if the case is opened. Bench Repair The secondary repair method is bench repair.
Theory of Operation and System Architecture Theory of Operation and System Architecture This section contains a high level overview of the theory of operation of the major functional modules of the monitor. Philips service providers perform Bench Repair at the assembly level. For this purpose, they should review this high level information. Please be aware that Philips only stocks assembly level parts. System Overview The monitor is a full function monitor for use on adult and pediatric patients.
Theory of Operation and System Architecture Figure 11-1 System Block Diagram 11-6 Training
Theory of Operation and System Architecture Isolated Front End The Isolated Front End section includes all of the circuitry to convert ECG, SpO2, and Temperature measurements to digital format; and to connect this information to the processor. The Respiration detection is obtained from two of the three electrodes of the ECG connections. NBP Front End The NBP section contains the pumps, valves, pressure measurement circuitry, and control circuitry for the non-invasive blood pressure measurement.
Theory of Operation and System Architecture The Contrast button operates in conjunction with the navigation wheel to determine the apparent contrast setting in the display. Changing the contrast is a change in the viewing angle. Outputs of the button and wheel are connected to the processor. Momentarily pressing the button sets the contrast to mid-range, factory-default value.
Theory of Operation and System Architecture Recorder The optional recorder module is installed in the right panel of the monitor. Refer to the C3 Instructions for Use guide for printing procedures. It provides users with the capability to obtain hard-copy records of selected vital sign information. Basic control of the recorder is implemented by two push-button controls on the recorder’s front panel.
Theory of Operation and System Architecture NBP Processing The NBP processing uses an oscillometric technique to provide needed measurements at selected intervals. This technique uses an inflatable sphygmomanometer cuff similar to those used by clinicians in routine measurements. A motorized pump inflates the cuff to approximately 180 mmHg initially, at which point the pressure effectively stops the flow of blood.
Troubleshooting Troubleshooting Service personnel concerned with troubleshooting should review the following topics in this manual: • Error Codes (see “Error Codes” on 6-3) • Boot Sequence (see “Performance Assurance Test” on 4-11) • Performance Verification (see “Performance Procedures” on 4-22) This is intended to be a practical, hands-on part of the training program. If possible, therefore, use a monitor to help you learn about troubleshooting the monitor.
Disassembly Audible alarm does not sound Speaker malfunction or speaker has a loose connector. Main PCB module malfunction. 1. Check speaker connector. 2. Replace speaker according to the instructions in “Replacing the Speaker” on 7-9. Replace the Main PCB module according to the instructions in “Removing the Main PCB Module” on 7-12. No sound from monitor Tone is switched off or low. No tone alarms Audible alarms are not switched on. Make sure audible alarms are switched on.
Monitor Applications and Algorithms Monitor Applications and Algorithms Information about... Found in... Supplies http://shop.medical.philips.
Monitor Applications and Algorithms 11-14 Training
12 Specifications This chapter includes all hardware, regulatory, and measuring specifications for the monitor.
Hardware Specifications Hardware Specifications Parameter Specification Size (excluding handle) Width 13.50 in (34.3 cm) Height 10.25 in (26.0 cm) Depth 7.50 in (19.0 cm) Weight (excluding accessories, options, cables) 14.0 lb (6.3 kg) Display (cold cathode fluorescent backlit) Screen Type TFT (color) Active Screen Size 8.3 x 6.2 in (210.8 x 157.5 mm): 10.4 in (264.2 mm) diagonally Resolution 640 x 480 pixels Recorder Type Thermal Weight 0.9 lb (.41 kg) Paper Width 50 mm Speeds 12.
Electrical Electrical Parameter Specification Power Sources Internal Battery 12 volt, 4.5 Ah, sealed, lead-acid Battery Operating Time 3 hours, fully charged battery at 25oC(without CO2) under normal (one NBP measurement every 15 minutes, all other modes active) conditions, 45 minutes if printing continuously, 1 hour, fully charged battery at 25oC (with CO2) under normal conditions, 30 minutes if printing continuously.
Measuring and Displaying Parameters Measuring and Displaying Parameters ECG 12-4 Specifications Parameter Specification Heart Rate Range 20-250 b/min Heart Rate Accuracy ±5 b/min Bandwidth: Normal Monitoring 0.5 Hz - 40 Hz Bandwidth: Extended Low Frequency Range (use selectable) 0.05 Hz - 40 Hz Leads 3 lead, user selectable Display Sweep Speeds 12.5, 25, and 50 mm/sec Pacemaker Detection Indicator on waveform display, user selectable ECG size (sensitivity) 0.
Measuring and Displaying Parameters ECG Standards Standards Meets the performance standards of ANSI/AAMI EC13-1992. Instead of a 1 mV standardizing voltage (section 3.2.2.9), a fixed, 1 cm reference bar is always present in the ECG display, along with the ECG size setting expressed in mV/cm. The following information references particular sections of ANSI/ AAMI EC13-1992. Respiration, leads-off sensing waveform. 3.1.2.1 (b) A 85 kHz square wave is used to inject 50 µA p-p differentially between RA and LL.
Measuring and Displaying Parameters Standards Pacemaker pulse rejection. 3.1.4.1, 3.1.4.2 With the exceptions noted below, the monitor will reject all single and double pacemaker pulses either 150 or 250 msec apart of amplitudes ±2 to ±700 mV from 0.1 to 2.0 msec, with and without under/overshoot.
Measuring and Displaying Parameters NBP (NonInvasive Blood Pressure) Parameter Specification Standards Meets performance standards of ANSI/ AAMI SP10-1992 Technique Oscillometric Cuff Pressure Display 10 to 300 mmHg Pulse Rate Range 40 to 200 b/min Blood Pressure Accuracy Mean error and standard deviation per ANSI/AAMI SP10-1992a Pulse Rate Accuracy Greater of ± 2 BMP or ± 2% Initial Cuff Inflation 180 mmHg Subsequence Cuff Inflation Prev SYS ±30 mmHg Blood Pressure Measurement Range Sys
Measuring and Displaying Parameters SpO2 Parameter Specification Range % Saturation 0 to 100% Pulse Rate 30 to 300 b/min Accuracy SpO2 0 - 69% ± not specified; M1191A, M1192A ± 2.5% (70 to 100%) M1194A ± 4.0% (70 to 100%) Accuracya with Philips disposable sensors (M1902B, M1903B, M1904B) and with Nellcor disposable sensors (OxiCliq A, P, N, I) 3% (70 to 100%). Pulse Rate 2% or 1 b/min (whichever is greater) Display Update Period 1 sec Calibration functional a.
Measuring and Displaying Parameters CO2 Complies with EN 864/ISO 9918 Parameter Specification Range 0 to 98 mmHg (0 to 13 kPa), or 13% CO2 whichever is lower. Flow Rate 50 ±7.5 ml/min Warm-up Time 180 seconds max.; typically 30 seconds Gas Sampling Delay Time 2.7 seconds Resolution Numeric 1.0 mmHg (0.1 kPa) Wave 0.1 mmHg (0.
Measuring and Displaying Parameters Trends Parameter Specification Type Graphical and tabular Memory Storage 12 hours, nonvolatile Data interval 20 seconds (stored data points is the average over 20-second interval) Graphical Trend Format One graph per vital sign Display range 2 hours, scrollable Vertical scaling 12-10 Specifications Heart Rate 50-100, 0-250, 25-125 b/min NBP 50-100, 0-150, 0-300 mmHg SpO2 0-100, 60-100, 80-100% CO2 0-100, 0-60, 20-50 mmHg Respiration Rate 0-20, 0-50