DL8765 EN-UK Specifications are subject to change without notice. © 2016 Koninklijke Philips N.V. All rights reserved Philips Consumer Lifestyle BV Tussendiepen 4, 9206AD Drachten, Netherlands Fax +31 (0)512594316 4222.100.4515.
1 1 2 3 4 15 14 5 13 6 7 12 8 9 11 10 16 17 English Introduction Congratulations on your purchase and welcome to Philips! To fully benefit from the support that Philips offers, register your product at www.philips.com/welcome.
Symbol for indoor use only. Symbol for 'Including RF transmitter'. This means that this device emits non-ionizing radiation. All devices with RF transmitters or that use RF electromagnetic energy must have a label with this symbol. Indicates caution.The user should consult the instructions for use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself. Symbol for micro-USB connector. Symbol for USB connector.
Caution - - - - - - - - - - - - - Always check the device before you use it. Do not use the device if it is damaged, as this may cause injury. The effectiveness of this blood pressure monitor has not been established in pregnant (including pre-eclamptic) women. Common arrhythmias (such as atrial or ventricular premature beats or atrial fibrillation) and peripheral artery disease / arteriosclerosis can affect the performance (accuracy) of this blood pressure monitor.
Display Sym bol Description Explanation Systolic blood pressure Maximum blood pressure. Also see section Systolic and diastolic pressure Diastolic blood pressure Minimum blood pressure, also see section Systolic and diastolic pressure. Heart rate Number of heartbeats per minute (pulse is typically equivalent to heart rate). Battery status Indicates status of battery during charging. Measure ment unit Measurement unit of blood pressure. Inflation indicator The cuff is inflating.
- Whether your blood pressure has reached a dangerous level. The following table shows the classification system for the blood pressure measurements used in this device. This system follows the classification system of the World Health Organisation (WHO).
Measuring blood pressure Tips for proper measurement - Rest for 5 minutes before you measure your blood pressure. - Wait at least 3 minutes between measurements. This allows your blood circulation to recover. - For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as directed by your doctor.
1 Activate the Philips HealthSuite health app and Bluetooth on your mobile device directly after a measurement. - Keep the mobile device and the blood pressure monitor at transmission distance (no more than 5 meters from each other, in the same room). 2 Once successfully connected, the measurement results are transmitted to the health app and the Bluetooth symbol lights up. - If the data transmission is successful, the measurement results are displayed in the health app.
Problem Possible cause Solution Fluctuations of blood pressure during the day are normal. Blood pressure fluctuates from minute to minute and normally shows a circadian rhythm over a 24-hour period, with highest readings in the afternoons and lowest readings at night. That is why, for comparable measurements, the measurements should be taken at approx. the same time of day. You are using The variations in medication. blood pressure can be greater if you are using medication.
Problem Possible cause Solution The blood pressure monitor and mobile device are more than 5 meters away from each other. Place your mobile device closer to the blood pressure monitor. You selected the wrong profile on the blood pressure monitor. Select the correct user profile on the blood pressure monitor before your measurement. Otherwise the data cannot be transmitted to your app. Repeat the measurement with the correct profile selected Specifications Power supply 3.
Emissions test Com pli ance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The device must emit electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey (a), should be less than the compliance level in each frequency range (b). Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.