User Manual

Problem Possible
cause
Solution
The blood
pressure
monitor and
mobile
device are
more than 5
meters away
from each
other.
Place your mobile
device closer to the
blood pressure
monitor.
You selected
the wrong
profile on the
blood
pressure
monitor.
Select the correct
user profile on the
blood pressure
monitor before
your measurement.
Otherwise the data
cannot be
transmitted to your
app. Repeat the
measurement with
the correct profile
selected
Specifications
Power supply
3.7V 420mAH built-in
rechargeable li-polymer
battery
Display Display with white
backlight
Visible area = 46.8mm (L)
x29.3mm (W)
Measurement method Oscillometric method
Measurement range Rated cuff pressure:
0kPa-40kPa
(0mmHg-300mmHg)
Measurement pressure:
4kPa-34kPa
(40mmHg-230mmHg)
heart rate: 40-199 beats
per minute
Accuracy Pressure:
5°C-40°C
within ±0.4kPa
(±3mmHg)
heart rate:±5% of
measurement result on
display
Normal operating
conditions
Temperature: 5°C to
40°C. Relative humidity:
85%RH. Atmospheric
pressure: 86kPa to
106kPa
Storage and
transportation
conditions
Temperature: -20°C to
60°C. Relative humidity:
10% to 93%. Atmospheric
pressure: 50kPa to
106kPa
Measurement
perimeter of the wrist
About 13.5cm-21.5cm
Net weight Approx. 100g
External dimensions Approx.
80mm×65mm×13.2mm
Accessories USB cable, user manual
Mode of operation Continuous operation
Degree of protection Type BF applied part
Protection against
ingress of water
IP22, This means:
protected against access
to hazardous parts with a
finger and against
vertically falling water
drops when tilted up to
15 degrees.
Device classification Battery Powered Mode:
Internally Powered ME
Equipment. Class II ME
Equipment
Caution: No modification of this equipment is
allowed.
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.
EMC Guidance
- The Blood Pressure Monitor needs special
precautions regarding EMC and needs to be
installed and put into service according to the
EMC information provided in the accompanying
documents.
- Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations,
walkie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
output power of 2 W yields d = 3.3 m at an
immunity level of 3V/m.
Table 1 Guidance and manufacturer's
declaration – electromagnetic emissions
- for all ME equipment and ME systems
Guidance and manufacturer’s declaration –
electromagnetic emissions
The device is intended for use in the
electromagnetic environment specified below. The
customer or the user of the device should assure
that it is used in such an environment.