DreamStation Go Heated Humidifier User manual Instructions for use
Table of contents 1. Safety information.............................................................................................................................. 1 2. System overview.................................................................................................................................3 3. Connecting and setting up the humidifier.....................................................................................5 Humidifier pop-up messages.......................................
1. Safety information Intended use The DreamStation Go Heated Humidifier is an accessory for the Philips Respironics DreamStation Go therapy devices to provide moisture to the patient circuit. It is intended for use in spontaneously breathing patients weighing over 30 kg (66 lbs), in the home or hospital/institutional environment, who use mask-applied positive pressure ventilation therapy. Warnings A warning indicates the possibility for injury to the user or the operator.
Cautions A caution indicates the possibility of damage to the device. • U.S. federal law restricts this device to sale by or on the order of a physician. • Do not place the humidifier directly onto carpet, fabric, or other flammable materials. • Do not place the device in or on any container that can collect or hold water. Take precautions to protect furniture from water damage. • Do not fill the water tank above the maximum fill line.
Symbols Glossary The following symbols may appear on the device. Refer to http://www.symbols.philips.com for a description of additional symbols used on this device and its packaging. Symbol Definition Symbol Maximum fill line IP22 Definition Drip Proof Equipment Protection against ingress of solid foreign objects ≥ 12.5 mm diameter.
System features and contents The figure below illustrates many of the device features and contents. In addition to the features shown below, your humidifier system may include a humidifier user manual.
3. Connecting and setting up the humidifier Connecting to the therapy device 1. Remove the humidifier beauty panel from the therapy device by pulling from either side of the panel. Pull here 2. Place the therapy device and heated humidifier (with an empty water tank) on a firm, flat surface. Caution Do not move the humidifier while the water tank has water in it. 3. Line up the front of the therapy device to the front of the humidifier.
Setting humidification mode and target time To access the humidifier settings in the Therapy Off menu, navigate to the My Comfort menu. From here you can view or adjust the following humidifier settings. Icon Text Description Humidification This displays the humidification mode being used. You can choose between Fixed or Adaptive humidification. Target Time This allows you to set the estimated amount of time you will use the device each night.
Humidifier pop-up messages Device pop-ups are messages that show up on the user interface screen. Condition Icon Description Possible Cause Action Humidification Error Humidifier error (only when humidifier is present). Humidifier heater plate error or humidifier not properly connected to therapy device. Turn off the device and disconnect it from power. Detach the humidifier, visually check that electrical contacts are clear, then reconnect humidifier and power cord.
4. Daily use Warning Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan. Cautions —— Do not attempt to fill the water tank while it is still inside the humidifier. —— Do not put any chemicals or additives into the water. Possible airway irritation or damage to the water tank may result.
4. Remove the tank lid by pulling up on the tank lid tab while holding the tank base. Pour out any remaining water and rinse the tank base. 5. Place the tank base on a firm, flat surface. Fill the tank base with water no higher than the maximum fill line . Note The water level may rise in the tank base when the tank lid is attached. 6. Reassemble the water tank. Align the tank lid with the tank base (1).
Disconnecting the humidifier Caution To avoid spilling, do not disconnect the humidifier from the therapy device with water in the tank. Remove the water tank from the humidifier before disconnecting the therapy device. 1. Disconnect power to the therapy device and tubing. 2. Remove the water tank from the humidifier. 3. Pick up the system. 4. Remove the humidifier by sliding the latch on the bottom of the humidifier, and pulling the humidifier away from the therapy device.
5. Cleaning for Home Use Warning Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan. Note Refer to the DreamStation Go therapy device user manual for adapter cleaning instructions. Cleaning the water tank base Hand washing can be performed daily. Dishwashing can be performed weekly.
2. Remove the baffle (2) by pressing in on the side tabs and pulling up on the baffle. 3. Wash the lid and components in a solution of warm water and a mild dishwashing detergent. Rinse with clean water. 4. Allow components to air dry. 5. Reassemble the baffle. Align the tabs on the baffle with the hooks on the tank lid. Press down until the tabs on each side of the baffle snap into place. 6. Reassemble the manifold. Align the manifold with the hooks on the tank lid.
6. Cleaning & Disinfection for Hospitals/Institutions If you are using the water tank base, water tank lid, and lid components on multiple patients, complete the following steps to clean and disinfect between each new user. Caution Only the hospital and institution cleaning and disinfection procedures listed in this manual are recommended by Respironics. Use of other cleaning and disinfecting processes, not specified by Respironics, may affect the performance of the product.
3. Wash the lid and components in a solution of warm water and a mild dishwashing detergent. Rinse with clean water. 4. Allow components to air dry. 5. Reassemble the baffle. Align the tabs on the baffle with the hooks on the tank lid. Press down until the tabs on each side of the baffle snap into place. 6. Reassemble the manifold. Align the manifold with the hooks on the tank lid. Gently press down on both manifold blocks until they snap into place. 7.
7. Maintenance instructions Warning Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan. Maintaining the water tank base If you use tap water in your humidifier water tank, follow the steps below to maintain the tank base. Perform these steps monthly or more frequently, as needed. 1.
8. Troubleshooting The table below lists some of the problems you may experience with your humidifier and possible solutions. Problem Why it happened What to do The humidifier and/ or components are cracked or damaged. The humidifier was dropped or mishandled. If the humidifier does not operate properly after being dropped or mishandled, contact your supplier or Philips Respironics. The therapy device is operating but the humidifier’s airflow is low or stopped.
Problem Why it happened What to do I’m having difficulty adjusting the heated humidifier setting. The blower is not turned on, or the humidifier is not fully connected. The humidifier setting can only be adjusted when therapy is turned ON. Confirm that the blower is turned on, and that the humidifier icon appears on the left side of the screen, then adjust to desired comfort. If the blower is on but the humidifier settings are not displayed on the Therapy ON screen, then unplug the device.
9. Additional notes Traveling with the system 1. Remove the water tank and empty all water and allow to air dry. 2. Put the empty water tank back into the device. 3. Pack your humidifier in your carry-on luggage. Caution Do not move the humidifier while the water tank has water in it. When you are traveling, the optional carrying case for your therapy device can be used for carry-on luggage only. The carrying case will not protect the humidifier if it is put through checked baggage.
Standards compliance This device is designed to conform to the following standards: • IEC 60601-1 General Requirements for Basic Safety and Essential Performance of Medical Electrical Equipment • IEC 60601-1-11 General Requirements for Basic Safety and Essential Performance in the Home Healthcare Environment • IEC 60601-1-6 General Requirements for Safety - Usability • IEC 62366 Application of Usability Engineering in Medical Devices • IEC 62304 Medical Device Software – Software Life-cycle Processes • ISO
Declared dual-number noise emissions values (in accordance with ISO 4871) Tube Size Sound Pressure Level (L) Uncertainty (K) Sound Power Level Uncertainty (K) 12 (mm) tubing type 33.9 dB(A) 2.0 db(A) 41.9 dB(A) 2.0 dB(A) 15 (mm) tubing type 33.3 dB(A) 2.0 db(A) 41.3 dB(A) 2.0 dB(A) 22 (mm) tubing type 33.2 dB(A) 2.0 db(A) 41.2 dB(A) 2.0 dB(A) Pressure accuracy Pressure Increments: 4-20cm H2O (in 0.
Limited Warranty Respironics, Inc., a Philips company, warrants that the system will be free from defects of workmanship and materials and will perform in accordance with the product specifications from the date of sale by Respironics, Inc. for a period of two (2) years for the therapy device and humidifier (not including the water tank) and a period of one (1) year for the battery pack (“Warranty Period”). If the product fails to perform in accordance with this limited warranty, Respironics, Inc.
1140901 Respironics Inc. 1001 Murry Ridge Lane Murrysville, PA 15668 USA Respironics Deutschland GmbH & Co.