Delivering a Synchronized Shock
M3500B HeartStart XLT Defibrillator/Monitor 7-3
7
Figure 7-1: Charging in Sync Mode
If desired, you may increase or decrease the selected energy level after
pressing
, by pressing repeatedly until the desired energy
level is displayed. The defibrillator charges to the modified energy level
automatically. Wait until the current charge reaches the selected energy
level before proceeding.
5. Make sure no one is touching the patient or anything connected to the
patient. Call out “Clear.” Then press and continue to hold
down until the shock is delivered.
The defibrillator shocks with the next detected R-wave.
WARNIN G Defibrillation current can cause operator or bystander injury. Do not touch the
patient, or equipment connected to the patient, during defibrillation.