User manual

Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specied
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
Immunity
test
IEC
60601
test
level
Com-
pli ance
level
Electromagnetic environment – guidance
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz
to 80
MHz
3 Vrms Portable and mobile RF communications equipment should
be used no closer to any part of the Philips Avent breast
pump, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.Recommended separation
distance d=1,17 P The breast pump can be turned o.
Radiated RF
IEC 61000-
4-3
3 V/m
80 MHz
to 2,5
Ghz
3 V/m 80 MHz to 800 Mhz d=1,17 P 800 MHz to 2,5 GHz d=2,33 P
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in metres (m).
Field strengths from xed RF transmitters, as determined
by an electromagnetic site survey, should be less than the
compliance level in each frequency range. Interference may
occur in the vicinity of equipment marked with the following
symbol: Caution: Although the breast pump is compliant
with applicable EMC directives, it may still be susceptible
to excessive emissions and/or may interfere with other
equipment. A consequence can be that the breast pump turns
o or the on/o button starts ashing (see troubleshooting for
more information), which is acceptable as this will not lead to
unacceptable risks. To prevent interference, keep other electric
equipment away from the breast pump during expressing.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by
absorption and reection from structures, objects and people.
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF
transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the
location in which the Philips Avent breast pump is used exceeds the applicable RF compliance level
above, the Philips Avent breast pump should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating
the Philips Avent breast pump.
English
26