User manual

10/ 2008
TESTFOX CC –1
12
The TESTFOX CC–1 has an automatic switch-over of measurement range.
Previous setting is therefore not required.
7.1 DC voltage range
The input resistance is 9 M.
Measuring range Resolution
Measurement accuracy
Overload protection
400,0 mV 100 µV ± (0,5 % of reading + 2 digits)
600 V
eff
600 V DC voltage
4,000 V 1 mV ± (0,5 % of reading + 2 digits)
600 V
eff
600 V DC voltage
40,00 V 10 mV ± (0,5 % of reading + 2 digits)
600 V
eff
600 V DC voltage
400,0 V 100 mV ± (0,5 % of reading + 2 digits)
600 V
eff
600 V DC voltage
600 V 1 V ± (0,5 % of reading + 2 digits)
600 V
eff
600 V DC voltage
7.2 AC voltage range
The input resistance is 9 MΩ parallel 100 pF. The measurement reading is
obtained by rectification of the average reading and displayed as the actual
value. Its calibration is adapted for a sinusoidal curve form. With deviations from
this form, the display value becomes less accurate. The additional error for the
following crest factors is as follows:
crest factor of 1.4 to 2.0: additional error ± 1.0%
crest factor of 2.0 to 2.5: additional error ± 2.5%
crest factor of 2.5 to 3.0: additional error ± 4.0%
Measuring range Resolution
Measurement accuracy
Overload protection
400,0 mV 100 µV
± (2,0 % of reading + 5 digits)*
in frequency range 50 Hz - 60 Hz
600 V
eff
600 V DC voltage
4,000 V 1 mV
± (1,5 % of reading + 5 digits)
in frequency range 40 Hz - 300 Hz
600 V
eff
600 V DC voltage
40,00 V 10 mV
± (1,5 % of reading + 5 digits)
in frequency range 40 Hz - 500 Hz
600 V
eff
600 V DC voltage
400,0 V 100 mV
± (1,5 % of reading + 5 digits)
in frequency range 40 Hz - 500 Hz
600 V
eff
600 V DC voltage
600 V 1 V
± (1,5 % of reading + 5 digits)
in frequency range 40 Hz - 500 Hz
600 V
eff
600 V DC voltage
* When input is opened, fluctuating up to 30 digits.
7.3 Resistance range
No-load voltage approx. 3 V, max. test current 0.1 mA
Measuring range Resolution
Measurement accuracy
Overload protection
400 Ω 100 mΩ
± (1,2 % of reading + 6 digits)
*1
600 V
eff
600 V DC voltage
4,000 kΩ 1 Ω
± (0,9 % of reading + 3 digit)
*2
600 V
eff
600 V DC voltage
40,00 kΩ 10
± (0,9 % of reading + 3 digit)
*2
600 V
eff
600 V DC voltage
400,0 kΩ 100 Ω
± (1,2 % of reading + 3 digit)
*2
600 V
eff
600 V DC voltage
4,000 MΩ 1 kΩ
± (1,2 % of reading + 3 digit)*
2
600 V
eff
600 V DC voltage
40,00 MΩ 10 kΩ
± (2,5 % of reading + 5 digit)
*1 *3
600 V
eff
600 V DC voltage
*1
in end area of display only, + 6 digits
*2
in end area of display only, + 3 digits
*3
maximum start-up time until display appears, 20 sec.
7.4 Continuity test
No-load voltage approx. 3 V, max. test current 0.7 mA
The built-in buzzer sounds when resistance is less than 50 Ω - 300 Ω.