Datasheet

https://www.phoenixcontact.com/us/products/2902851
04/20/2016Page3/6
Thermocouple measuring transducer-MINI MCR-TC-UI-NC-2902851
Technical data
Connection data
Conductor cross section flexible max. 2.5 mm²
Stripping length 12 mm
Screw thread M3
General
Transmission error thermocouples
0.1 % * 600 K / set measuring range; 0.1 % > 600 K (E, J, K, N, T, L, U, M
Gost, L Gost)
0.2 % * 600 K / set measuring range; 0.2 % > 600 K (B, R, S, A1, A2, A3)
0.2%* 600K / set measuring range; 0.2%> 600K (E, J, K, N, T, L, U, M
Gost, L Gost); Highspeed Mode
0.4%* 600K / set measuring range; 0.4%> 600K (B, R, S, A1, A2, A3);
Highspeed Mode
Maximum temperature coefficient ≤ 0.01 %/K
Cold point error, max. < 3 K
Typical cold point errors < 2 K
Status display LED red
Protective circuit Transient protection
Electrical isolation Basic insulation according to EN 61010
Overvoltage category II
Degree of pollution 2
Rated insulation voltage 50 V AC/DC
Test voltage, input/output/supply 1.5 kV (50 Hz, 1 min.)
Electromagnetic compatibility Conformance with EMC Directive 2004/108/EC
Noise emission EN 61000-6-4
Noise immunity
EN 61000-6-2 When being exposed to interference, there may be minimal
deviations.
Color green
Housing material PBT
Mounting position any
Conformance CE-compliant
ATEX  II 3 G Ex nA IIC T4 Gc X
UL, USA / Canada UL 508 Listed
Class I, Div. 2, Groups A, B, C, D T4
Class I, Zone 2, Group IIC
Certificate of classification DNV GL 14085-15HH
Standards and Regulations
Electromagnetic compatibility Conformance with EMC Directive 2004/108/EC
Noise emission EN 61000-6-4
Designation Electromagnetic RF field