6 Data Management
Data Management
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 6-13
©2006-2013 Physio-Control, Inc.
Figure 6-3 IrDA Connections
DATA TRANSFER FROM TrueCPR DEVICE
Note: The TrueCPR device may not be available in all countries. Contact your local
Physio-Control representative for more information.
You can transfer data from the Physio-Control TrueCPR device to Physio-Control post-event
review tools using the TrueCPR device port on the CodeManagement Module. To transfer data:
1 Confirm the LIFEPAK 20e defibrillator/monitor is either turned off or in archives mode.
2 Connect an approved USB cable between the TrueCPR device and the CodeManagement
Module.
Note: Use only a USB cable that meets the following specifications: USB 2.0 A-Male to
Mini-B 5-pin Male, 28/24 AWB cable with ferrite core (gold plated), 1 m (3 ft).
3 Turn on the TrueCPR device and follow the instructions in the TrueCPR device Instructions
for Use.
Notes:
It may take up to one minute for the TrueCPR device to display the Data Transfer screen.
Data transfer takes approximately 3–5 minutes.
If you attempt to transmit data from the LIFEPAK 20e defibrillator/monitor and the TrueCPR
device simultaneously, the data from the defibrillator will be transmitted first.
IrDA
Adapter/
Computer
Maximum Distance: 1.0 m (3.28 feet)
Defibrillator