6-14 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
TROUBLESHOOTING TIPS FOR DATA TRANSMISSION
Table 6-4 Troubleshooting Tips for Data Transmission
Observation Possible Cause Corrective Action
1 SEND DATA option does
not appear in Options/
Archives screen.
CodeManagement Module not
properly connected to
defibrillator.
Contact qualified service
personnel.
2
TRANSMISSION FAILED
message appears.
No wireless connection to
LIFENET System.
Verify Wireless switch on
side of CodeManagement
Module is in the
ON position.
See "Side View of
CodeManagement Module,"
page 2-15.
Contact your facility’s
LIFENET System
administrator to verify the
receiving computer is
operational.