7-10 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
GENERAL TROUBLESHOOTING TIPS
If a problem with the defibrillator/monitor is detected during operation or testing, refer to the
troubleshooting tips in Table 7-2. If the problem cannot be corrected, remove the defibrillator/
monitor from use and contact qualified service personnel.
Table 7-2 General Troubleshooting Tips
Observation Possible Cause Corrective Action
1 No power when
defibrillator/monitor is
turned
ON.
Low battery voltage. Connect to AC power.
2 Defibrillator/monitor
operates, but screen is
blank.
Operating temperature is
too low or too high.
Connect to AC power.
Screen is not displaying
properly.
Contact qualified service personnel.
3
CHECK PRINTER
message appears.
Printer paper jams, slips, or
misfeeds.
Printer is out of paper.
Reinstall paper.
Add new paper.
If problem persists, contact qualified
service
personnel.
4 AC Mains LED on
defibrillator/monitor not
illuminated when
connected to AC power.
Loose or improper
connection between
defibrillator and power
source.
Check power connections and
cables.
Loose or improper
connection between
defibrillator and
CodeManagement Module.
Contact qualified service personnel.
5 Service LED
illuminates.
Device self-test circuitry
detects service condition.
Continue to use defibrillator or
pacemaker if needed.
Turn device off and then on again.
Note that this creates a new Patient
Record.
If service LED does not clear,
remove device from active use.
Report occurrence of service LED to
qualified service personnel.
6 ECG monitoring
problems.
Refer to Section 3, page 3-5.
7 Problems with AED
operation.
Refer to Section 4, page 4-13.
8 Problems with
defibrillation/
synchronized
cardioversion.
Refer to Section 4, page 4-22.
9 Problems with pacing. Refer to Section 4, page 4-25.
10 Displayed time is
incorrect.
Time is incorrectly set. Change the time setting. Refer to
Section 2, page 2-7.