User Manual

Defining Setup Options
8-8 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
CPR METRONOME SETUP MENU
The CPR Metronome Setup menu allows you to enable the CPR metronome for AED mode, and
define the C:V (compression to ventilation) ratio for each Age-Airway setting. The C:V ratio is
adjustable to provide different settings based on patient age and airway status and according to
local medical protocols.
PRESHOCK
CPR
Set CPR interval after shock
advised decisions
OFF, 15, 30 SECONDS
Table 8-7 AED Mode Energy Protocol Setup Menu
Menu Item Help Message Options
PRESET
PROTOCOLS
Select a preset energy
protocol
Energy 1: 150, 175, 200, 225, 250, 275,
300
, 325, 360
Energy 2: 150, 175, 200, 225, 250, 275,
300
, 325, 360
Energy 3: 150, 175, 200, 225, 250, 275,
300
, 325, 360
Energy 2 cannot be less than Energy 1.
Energy 3 cannot be less than Energy 2.
FLEXIBLE
PROTOCOLS
Repeat previous energy after
NO SHOCK ADVISED
ON
or OFF.
STACKED
SHOCKS
Enable consecutive shocks
without CPR
ON or OFF.
Table 8-8 CPR Metronome Setup Menu
Menu Item Help Message Options
METRONOME
Enable metronome during
CPR
ON or OFF
(Turns metronome on or off for AED mode.)
ADULT – NO
AIRWAY
Set Adult – No Airway
compression/ventilation ratio
30:2, 16:1, 15:2, 12:1, 10:1, 100:0
ADULT –
AIRWAY
Set Adult – Airway
compression/ventilation ratio
30:2, 16:1, 15:2, 12:1, 10:1, 100:0
YOUTH – NO
AIRWAY
Set Youth – No Airway
compression/ventilation ratio
30:2, 16:1, 15:2, 12:1, 10:1, 100:0
YOUTH –
AIRWAY
Set Youth – Airway
compression/ventilation ratio
30:2, 16:1, 15:2, 12:1, 10:1, 100:0
Table 8-6 AED Mode CPR Setup Mode (Continued)
Menu Item Help Message Options