8 Defining Setup Options
Defining Setup Options
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 8-13
©2006-2013 Physio-Control, Inc.
RESET DEFAULTS SETUP MENU
Use the Reset Defaults menu to configure the defibrillator for all factory default settings.
PRINT DEFAULTS
Use the Print Defaults menu to print the current device configuration.
SEND CONFIGURATION SETUP MENU
Use the Send Configuration Setup menu to transfer the setup in one device to overwrite the
setup in another device. You can send setup configurations between devices with different
features because all devices have identical setup menus, regardless of features.
Note: If a CodeManagement Module is attached to the defibrillator, the Send Configuration
Setup option will not function. In this case, you should use the LIFENET System to manage
device configurations.
To send the setup from one device to another:
1 Connect the transport configuration cable to the system connector port on both devices.
2 Turn on the sending defibrillator and enter setup mode (refer to page 8-2).
3 Select the
SEND CONFIG... menu option.
4 Turn on the receiving defibrillator.
5 On the sending defibrillator, click
SEND and then follow the screen prompts.
Table 8-19 Reset Defaults Setup Menu
Menu Item Help Message Options
CANCEL
Cancel and return to Setup
screen
Cancels reset operation.
RESET
Reset to factory configuration
settings
Resets setup parameters to the factory
default settings, except the maintenance
interval, which remains unchanged.
Table 8-20 Send Configuration Setup Menu
Menu Item Help Message Options
SEND
Send device configuration to
another device.
Connect devices with a cable, display this
screen on both defibrillators, and then
select
SEND.
PREVIOUS
PAGE
Go back to previous page. Cancels the operation.