A-4 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
CO2
CO2 Range
0 to 99 mmHg (0 to 13.2 kPa)
Units: mmHg, %, or kPa
CO2 Accuracy CO2 partial pressure at sea
level:
Accuracy:
(0–80 bpm)*
0 to 38 mmHg
(0 to 5.1 kPa)
±2 mmHg
(0.27 kPa)
39 to 99 mmHg
(5.2 to 13.2 kPa)
±5% of reading + 0.8% for
every 1 mmHg (0.13 kPa)
above 38 mmHg (5.1 kPa)
(>80 bpm)* 0 to 18 mmHg
(0 to 2.4 kPa)
±2 mmHg
(0.27 kPa)
19 to 99 mmHg
(2.53 to 13.2 kPa)
±4 mmHg (0.54 kPa) or
±12% of reading, whichever
is higher
*For RR > 60 bpm, to achieve specified CO2 accuracy, the
Microstream
®
Filterline
®
H Set for infants must be used.
Drift of Measurement
Accuracy
No drift in accuracy for at least 6 hours
Respiration Rate
Accuracy
0 to 70 bpm: ±1 bpm
71 to 99 bpm: ±2 bpm
Respiration Rate
Range
0 to 99 breaths/minute
Flow Rate 42.5 to 65 ml/min (measured by volume)
Rise Time 190 msec
Response Time 4.5 seconds, maximum (includes delay time and rise time)
Initialization Time 30 seconds (typical), 10–180 seconds
Ambient Pressure Automatically compensated internally
Waveform Scale
Factors
Autoscale, 0–20 mmHg (0–4 Vol%), 0–50 mmHg (0–7 Vol%),
0–100 mmHg (0–14 Vol%)
PRINTER
Prints continuous strips of the displayed patient information
Paper Size 50 mm (2.0 in.)
Print Speed
Continuous ECG 25 mm/sec
±5% (measured in accordance
with AAMI EC-11 (1991), 4.2.5.2)
25 mm/sec print speed for CODE SUMMARY Reports
Delay 8 seconds
Autoprint Waveform events print automatically (user configurable)