User Manual

Appendix C
Screen Messages
LIFEPAK 20e Defibrillator/Monitor Operating Instructions C-3
©2006-2013 Physio-Control, Inc.
LL LEADS OFF
F LEADS OFF
ECG electrode LL is disconnected.
ECG electrode F is disconnected.
LOW BATTERY
Battery status indicator shows one yellow segment; low battery
condition exists.
LOW BATTERY:
CONNECT TO AC
POWER
Battery status indicator shows one flashing red segment; very low
battery condition exists.
LOW IMPEDANCE–
RECHARGING
Patient impedance of <15 ohms detected.
MOTION DETECTED!/
STOP MOTION!
The defibrillator detected motion during ECG analysis, thereby
inhibiting analysis.
NO SHOCK ADVISED
The defibrillator does not detect a shockable rhythm.
NON-DEMAND
The pacemaker is in nondemand (asynchronous) mode.
PACER FAULT
The pacemaker detects a pacing fault condition due to high pacing
rate or loss of interprocessor communication. Pacing function stops.
PACING STOPPED
Pacing stops and this message appears whenever any of the following
occur: pacer electrodes off, pacer cable disconnected, or pacer failure
due to high pacing rate or high impedance.
PAUSED
The pacing PAUSE button was pressed and held. Current pulses are
applied at reduced frequency while the mA and ppm settings are
maintained.
PUSH ANALYZE
Press ANALYZE to begin ECG analysis.
PUSH AND HOLD
PADDLE BUTTON TO
SHOCK!
The front panel SHOCK button is disabled if internal paddles are
attached. This message appears if the defibrillator is in Sync mode
and you attempt to transfer energy by pressing the front panel
SHOCK
button.
PUSH AND HOLD
PADDLE BUTTONS TO
SHOCK!
The front panel SHOCK button is disabled if standard paddles are
attached. This message appears if the defibrillator is in Sync mode
and you attempt to transfer energy by pressing the front panel
SHOCK
button.
PUSH AND HOLD
SHOCK BUTTON!
The defibrillator is in Sync mode, fully charged, and ready to provide
therapy.
PUSH AND HOLD
SHOCK BUTTONS!
The defibrillator is in Sync mode, fully charged, and ready to provide
therapy with standard paddles.
PUSH PADDLE
BUTTON TO SHOCK!
The front panel SHOCK button is disabled if internal paddles are
attached. This message appears if you attempt to transfer energy by
pressing the front panel
SHOCK button.
Table C-1 Summary of Screen Messages (Continued)
Message Description