User Manual

Monitoring
3-8 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Loose connection.
Damaged cable or connector/
lead wire.
Check/reconnect cable
connections.
Inspect ECG and therapy cables.
Replace if damaged.
Check cable with simulator and
replace if malfunction observed.
Misplaced electrodes/lead
wire.
Confirm correct placement.
Select lead view with optimal QRS
detection.
Noise because of radio
frequency interference (RFI).
Check for equipment causing RFI
(such as a radio transmitter) and
relocate or turn off equipment
power.
4 Baseline wander
(low frequency/high
amplitude artifact).
Inadequate skin preparation.
Poor electrode-skin contact.
Diagnostic frequency
response.
Prepare skin and apply new
electrodes.
Check electrodes for proper
adhesion.
Print ECG in monitor frequency
response.
5 Fine baseline artifact
(high frequency/low
amplitude).
Inadequate skin preparation.
Isometric muscle tension in
arms/legs.
Prepare skin and apply new
electrodes.
Confirm that limbs are resting on a
supportive surface.
Check electrodes for proper
adhesion.
6 Systole beeps not
heard or do not occur
with each QRS
complex.
Volume too low.
QRS amplitude too small to
detect.
•Adjust volume.
Change ECG lead.
7 Monitor displays
dashed lines with no
ECG leads off
messages.
PADDLES lead selected but
patient connected to ECG
cable.
Select one of the limb leads.
8 Heart rate (HR)
display different than
pulse rate.
Monitor is detecting the
patient’s internal pacemaker
pulses.
Prepare skin and apply new
electrodes in different location.
Select lead view with optimal QRS
detection.
9 Internal pacemaker
pulses difficult to see.
Pulses from pacemaker are
very small. Monitor the visibility
of frequency response limits.
Turn on internal pacemaker
detector (refer to page 3-6).
Connect ECG cable and select
ECG lead instead of paddles.
Print ECG in diagnostic mode
(refer to page 3-5).
Table 3-2 Troubleshooting Tips for ECG Monitoring (Continued)
Observation Possible Cause Corrective Action