User Manual

Monitoring
3-14 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Troubleshooting Tips for SpO2
Table 3-3 Troubleshooting Tips for SpO2
Observation Possible Cause Corrective Action
1 The oximeter measures a
pulse, but there is no oxygen
saturation or pulse rate.
Excessive patient motion. Keep patient still.
Check that sensor is secure.
Relocate sensor.
Replace sensor.
Patient perfusion may be too
low.
Check patient.
Increase sensitivity.
2
SpO2 or pulse rate changes
rapidly; pulse amplitude is
erratic.
Excessive patient motion. Keep patient still.
Check that sensor is secure.
Relocate sensor.
Replace sensor.
Increase sensitivity.
An electrosurgical unit (ESU)
may be interfering with
performance.
Move the monitor as far as
possible from the ESU.
Plug the ESU and monitor
into different circuits.
Move the ESU ground pad
as close to the surgical site
as possible.
Sensor may be damp,
replace it.
Remove sensor extension
cable and connect the
sensor directly.
3 No
SpO2 value is displayed. Oximeter may be performing a
self-test (requires 10 seconds).
Wait for completion.
4
SPO2: NO SENSOR
DETECTED
message
appears.
Sensor is not connected to
patient or cable disconnects
from device.
Check that sensor and
cable are connected
properly.
5
SPO2: CHECK SENSOR
message appears.
Sensor is disconnected from
patient or cable.
Attach the sensor.
Check that sensor is secure.
Excessive ambient light. Remove or block light
source if possible.
Cover sensor with opaque
material, if necessary.
Patient has a weak pulse or
low blood pressure.
Test sensor on someone
else.
Check if patient perfusion is
adequate for sensor
location.
Check if sensor is secure
and not too tight.
Check that sensor is not on
extremity with blood
pressure cuff or
intravascular line.
Change sensor location.