3 Monitoring
Monitoring
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 3-15
©2006-2013 Physio-Control, Inc.
6 SPO2: UNKNOWN SENSOR
message appears.
A sensor is connected to the
device that is not compatible
with the Masimo Sp02 module.
Refer to "Pulse Oximeter
Sensors," page 3-13 for
sensor compatibility.
7
SPO2: SEARCHING FOR
PULSE
message appears.
A sensor is connected to the
patient and is searching for a
pulse.
Wait for completion.
8
SPO2: LOW PERFUSION
message appears.
Patient has a weak pulse. Change sensor location.
Table 3-3 Troubleshooting Tips for SpO2 (Continued)
Observation Possible Cause Corrective Action